Inflammatory tissue reactions around aseptically loose cemented hip prostheses: A retrieval study of the Spectron EF stem with Reflection All‐Poly acetabular cup. Issue 7 (31st January 2022)
- Record Type:
- Journal Article
- Title:
- Inflammatory tissue reactions around aseptically loose cemented hip prostheses: A retrieval study of the Spectron EF stem with Reflection All‐Poly acetabular cup. Issue 7 (31st January 2022)
- Main Title:
- Inflammatory tissue reactions around aseptically loose cemented hip prostheses: A retrieval study of the Spectron EF stem with Reflection All‐Poly acetabular cup
- Authors:
- Wolf, Susann
Johannessen, Anne Christine
Ellison, Peter
Furnes, Ove
Hallan, Geir
Rogg, Katharina
Skarstein, Kathrine
Høl, Paul Johan - Abstract:
- Abstract: The cemented Spectron EF stem in combination with the cemented non‐crosslinked Reflection All‐Poly cup showed a high rate of mid‐term aseptic loosening. However, the failure mechanisms are not fully known. We assessed the inflammatory tissue reactions and wear particles in periprosthetic tissues, implant wear and blood metal ion levels in 28 patients with failed implants. Histological analysis showed a macrophage pre‐dominant pattern with randomly distributed lymphocytes, with various amounts of neutrophils and giant cells. The number of different cell types in the tissue samples from patients in the cup group and in the stem group was similar. Wear particles, mainly ZrO2, CoCrMo, and polyethylene particles of different sizes and shapes, were associated with macrophages/giant cells, and total particle load/mm 2 was higher in cases of stem loosening. The Spectron EF stems were heavily worn, abraded, and polished. Stem abrasion correlated with metal ion concentrations in blood. The median polyethylene wear rate of the Reflection cups was 0.23 mm/year. The high proximal roughness of the Spectron EF stem resulted in excessive cement wear during loosening. The resulting inflammatory tissue responses to the degradation products both from the cup and the stem led to massive osteolysis and subsequent implant loosening.
- Is Part Of:
- Journal of biomedical materials research. Volume 110:Issue 7(2022)
- Journal:
- Journal of biomedical materials research
- Issue:
- Volume 110:Issue 7(2022)
- Issue Display:
- Volume 110, Issue 7 (2022)
- Year:
- 2022
- Volume:
- 110
- Issue:
- 7
- Issue Sort Value:
- 2022-0110-0007-0000
- Page Start:
- 1624
- Page End:
- 1636
- Publication Date:
- 2022-01-31
- Subjects:
- adverse local tissue reactions -- blood metal ions -- cemented total hip arthroplasty -- inflammation -- osteolysis -- wear debris
Biomedical materials -- Periodicals
610.28 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/jbm.b.35023 ↗
- Languages:
- English
- ISSNs:
- 1552-4973
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4953.725000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 21472.xml