Device to predict pelvic floor integrity during vaginal delivery: an intra- and interrater reliability study of the Epi-no distensibility measurement. (2nd November 2021)
- Record Type:
- Journal Article
- Title:
- Device to predict pelvic floor integrity during vaginal delivery: an intra- and interrater reliability study of the Epi-no distensibility measurement. (2nd November 2021)
- Main Title:
- Device to predict pelvic floor integrity during vaginal delivery: an intra- and interrater reliability study of the Epi-no distensibility measurement
- Authors:
- Paschoal, Aline
Uchiyama Nakamara, Mary
Araujo Júnior, Edward
Petricelli, Carla Dellabarba
Alexandre, Sandra Maria
Zanetti, Miriam Raquel Diniz - Abstract:
- Abstract: Objective: To determine the test–retest reliability of the pelvic floor distensibility evaluation with Epi-no ® device. Methods: Prospective, blinded observational study with 32 women pregnant from 35 gestational weeks. An Epi-no ® balloon was inflated within the pregnant woman's vagina and the maximum circumference achieved was measured using a standard metric measuring tape. Measurements were performed twice in a day by two physiotherapists in a randomized order. Intraobserver reproducibility was obtained with another evaluation 7–14 d after the initial examination. The intraclass correlation coefficient (ICC) was used to determine the intra and interrater reliability and the respective 95% confidence intervals with an alpha level of 0.05. Results: From a total of 32 included pregnant women, 28 were analyzed; they were 29.3 (±5.97) years old, were at 37 (±1.3) weeks' gestation and presented a mean Epi-no ® circumference of 20.4 (±2.4) cm. Regarding the intraobserver analysis, examiner 1 observed a maximum circumference of 20.6 (±2.6) cm in the first evaluation and 20.2 (±2.4) cm in the second evaluation, with good reliability (ICC = 0.85). Examiner 2 observed 19.7 (±2.4) cm in the first evaluation and 21.0 (±2.0) cm in the second one, with moderate reliability (ICC = 0.58). Regarding the interobserver analysis, there was good reliability on two evaluation days, with an ICC of 0.76 and 0.82, respectively. Conclusions: Epi-no ® device is a reliable device forAbstract: Objective: To determine the test–retest reliability of the pelvic floor distensibility evaluation with Epi-no ® device. Methods: Prospective, blinded observational study with 32 women pregnant from 35 gestational weeks. An Epi-no ® balloon was inflated within the pregnant woman's vagina and the maximum circumference achieved was measured using a standard metric measuring tape. Measurements were performed twice in a day by two physiotherapists in a randomized order. Intraobserver reproducibility was obtained with another evaluation 7–14 d after the initial examination. The intraclass correlation coefficient (ICC) was used to determine the intra and interrater reliability and the respective 95% confidence intervals with an alpha level of 0.05. Results: From a total of 32 included pregnant women, 28 were analyzed; they were 29.3 (±5.97) years old, were at 37 (±1.3) weeks' gestation and presented a mean Epi-no ® circumference of 20.4 (±2.4) cm. Regarding the intraobserver analysis, examiner 1 observed a maximum circumference of 20.6 (±2.6) cm in the first evaluation and 20.2 (±2.4) cm in the second evaluation, with good reliability (ICC = 0.85). Examiner 2 observed 19.7 (±2.4) cm in the first evaluation and 21.0 (±2.0) cm in the second one, with moderate reliability (ICC = 0.58). Regarding the interobserver analysis, there was good reliability on two evaluation days, with an ICC of 0.76 and 0.82, respectively. Conclusions: Epi-no ® device is a reliable device for physiotherapists to measure pelvic floor distention during pregnancy. … (more)
- Is Part Of:
- Journal of maternal-fetal & neonatal medicine. Volume 34:Number 21(2021)
- Journal:
- Journal of maternal-fetal & neonatal medicine
- Issue:
- Volume 34:Number 21(2021)
- Issue Display:
- Volume 34, Issue 21 (2021)
- Year:
- 2021
- Volume:
- 34
- Issue:
- 21
- Issue Sort Value:
- 2021-0034-0021-0000
- Page Start:
- 3481
- Page End:
- 3487
- Publication Date:
- 2021-11-02
- Subjects:
- Epi-no -- pelvic floor -- physical therapy -- pregnant women -- reproducibility
Obstetrics -- Periodicals
Perinatology -- Periodicals
Infants (Newborn) -- Diseases -- Periodicals
Neonatology -- Periodicals
618.2 - Journal URLs:
- http://informahealthcare.com/loi/jmf ↗
http://informahealthcare.com ↗ - DOI:
- 10.1080/14767058.2019.1685970 ↗
- Languages:
- English
- ISSNs:
- 1476-7058
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5012.332000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 21367.xml