A LC-MS3 strategy to determine lamotrigine by Q-Q-trap tandem mass spectrometry coupled with triple stage fragmentation to enhance sensitivity and selectivity. Issue 38 (10th September 2021)
- Record Type:
- Journal Article
- Title:
- A LC-MS3 strategy to determine lamotrigine by Q-Q-trap tandem mass spectrometry coupled with triple stage fragmentation to enhance sensitivity and selectivity. Issue 38 (10th September 2021)
- Main Title:
- A LC-MS3 strategy to determine lamotrigine by Q-Q-trap tandem mass spectrometry coupled with triple stage fragmentation to enhance sensitivity and selectivity
- Authors:
- Shi, Meiyun
Jiang, Qiuhong
Lyu, Qiushi
Yuan, Zhengting
Deng, Lili
Yin, Lei - Abstract:
- Abstract : The novelty of this study is the MS 3 technique resulting in high selectivity and high sensitivity. Abstract : A high-performance liquid chromatography tandem mass spectrometry cubed (HPLC/MS 3 ) method was developed and validated to quantify lamotrigine in human plasma with carbamazepine as an internal standard. The HPLC/MS/MS system is composed of a Shimadzu UFLC XR high-performance liquid chromatograph coupled with a hybrid linear ion trap triple quadrupole mass spectrometer. Following simple protein precipitation with methanol, the separation of lamotrigine and carbamazepine was performed on an Agilent Poroshell 120 SB-C18 column (4.6 × 50 mm, 2.7 μm) using gradient elution with 0.1% formic acid in water (solvent I) and 0.1% formic acid in methanol (solvent II) at a flow rate of 0.8 mL min −1 . The total run time for each sample was 5 min. The method was validated for accuracy, precision, linearity, lower limit of quantification (LLOQ), selectivity, and other parameters. The LC/MS 3 method was linear in the concentration range of 0.50–50.0 μg mL −1 ( R 2 ≥ 0.995). The LLOQ was 0.5 μg mL −1, requiring only 30 μL of human plasma. Intra- and inter-day accuracies were <6.17% and precisions were <11.4% at all concentrations. The absolute recoveries (%) and matrix effect (%) for lamotrigine in human plasma were between 83.8 and 90.7. The developed and validated LC-MS 3 assay was successfully applied to monitor the lamotrigine levels in human plasma after theAbstract : The novelty of this study is the MS 3 technique resulting in high selectivity and high sensitivity. Abstract : A high-performance liquid chromatography tandem mass spectrometry cubed (HPLC/MS 3 ) method was developed and validated to quantify lamotrigine in human plasma with carbamazepine as an internal standard. The HPLC/MS/MS system is composed of a Shimadzu UFLC XR high-performance liquid chromatograph coupled with a hybrid linear ion trap triple quadrupole mass spectrometer. Following simple protein precipitation with methanol, the separation of lamotrigine and carbamazepine was performed on an Agilent Poroshell 120 SB-C18 column (4.6 × 50 mm, 2.7 μm) using gradient elution with 0.1% formic acid in water (solvent I) and 0.1% formic acid in methanol (solvent II) at a flow rate of 0.8 mL min −1 . The total run time for each sample was 5 min. The method was validated for accuracy, precision, linearity, lower limit of quantification (LLOQ), selectivity, and other parameters. The LC/MS 3 method was linear in the concentration range of 0.50–50.0 μg mL −1 ( R 2 ≥ 0.995). The LLOQ was 0.5 μg mL −1, requiring only 30 μL of human plasma. Intra- and inter-day accuracies were <6.17% and precisions were <11.4% at all concentrations. The absolute recoveries (%) and matrix effect (%) for lamotrigine in human plasma were between 83.8 and 90.7. The developed and validated LC-MS 3 assay was successfully applied to monitor the lamotrigine levels in human plasma after the administration of lamotrigine. … (more)
- Is Part Of:
- Analytical methods. Volume 13:Issue 38(2021)
- Journal:
- Analytical methods
- Issue:
- Volume 13:Issue 38(2021)
- Issue Display:
- Volume 13, Issue 38 (2021)
- Year:
- 2021
- Volume:
- 13
- Issue:
- 38
- Issue Sort Value:
- 2021-0013-0038-0000
- Page Start:
- 4478
- Page End:
- 4484
- Publication Date:
- 2021-09-10
- Subjects:
- Chemistry, Analytic -- Periodicals
Analytical biochemistry -- Periodicals
Chemical laboratories -- Standards -- Periodicals
543.1905 - Journal URLs:
- http://pubs.rsc.org/en/Journals/JournalIssues/AY ↗
http://www.rsc.org/ ↗ - DOI:
- 10.1039/d1ay01372f ↗
- Languages:
- English
- ISSNs:
- 1759-9660
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0897.103700
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 21398.xml