Bag valve resuscitator versus Mapleson C circuit during manual ventilation: A bench top study. (June 2022)
- Record Type:
- Journal Article
- Title:
- Bag valve resuscitator versus Mapleson C circuit during manual ventilation: A bench top study. (June 2022)
- Main Title:
- Bag valve resuscitator versus Mapleson C circuit during manual ventilation: A bench top study
- Authors:
- Giani, Marco
Lucchini, Alberto
Moretto, Lorenza
Di Pierro, Michela
Lo Re, Fabio
Mancini, Paolo Giobbi
Palano, Salvatore
Foti, Giuseppe
Bellani, Giacomo
Bronco, Alfio - Abstract:
- Abstract: Background: Manual ventilation is life saving in critically ill patients. The lack of airway pressure monitoring makes it operator and device dependent. In this bench top-study, we compared a self- inflating bag valve resuscitator and a Mapleson C circuit during manual ventilation performed by critical care nurses under normal and pathologic conditions, with a special focus on delivered positive end expiratory pressure (PEEP). Methods: Three different respiratory patterns (normal, restrictive and obstructive) were reproduced by a breathing simulator. Twenty nurses provided manual ventilation with a specific ventilatory pattern. Airway pressure, tidal volume and respiratory rate were recorded. Absolute value, error (difference between recorded and target values) and variability of PEEP were analysed. Results: 3820 breathing traces were analysed. PEEP error was significantly higher with Mapelson C (43.3% vs 5.9% respectively, p < 0.001). This finding was confirmed regardless of operator skill and scenario. PEEP was more variable with Mapelson C (p < 0.05 in all scenarios). Ventilation of obstructive patients with Mapelson C resulted in higher PEEP levels compared to the reference value. Conversely, in the restrictive setting, PEEP was lower. Difference between PEEP and the minimum pressure recorded during the respiratory cycle was significantly higher with Mapelson C (p < 0.05). Conclusions: Manual ventilation with a Mapleson C circuit delivered a less accurate andAbstract: Background: Manual ventilation is life saving in critically ill patients. The lack of airway pressure monitoring makes it operator and device dependent. In this bench top-study, we compared a self- inflating bag valve resuscitator and a Mapleson C circuit during manual ventilation performed by critical care nurses under normal and pathologic conditions, with a special focus on delivered positive end expiratory pressure (PEEP). Methods: Three different respiratory patterns (normal, restrictive and obstructive) were reproduced by a breathing simulator. Twenty nurses provided manual ventilation with a specific ventilatory pattern. Airway pressure, tidal volume and respiratory rate were recorded. Absolute value, error (difference between recorded and target values) and variability of PEEP were analysed. Results: 3820 breathing traces were analysed. PEEP error was significantly higher with Mapelson C (43.3% vs 5.9% respectively, p < 0.001). This finding was confirmed regardless of operator skill and scenario. PEEP was more variable with Mapelson C (p < 0.05 in all scenarios). Ventilation of obstructive patients with Mapelson C resulted in higher PEEP levels compared to the reference value. Conversely, in the restrictive setting, PEEP was lower. Difference between PEEP and the minimum pressure recorded during the respiratory cycle was significantly higher with Mapelson C (p < 0.05). Conclusions: Manual ventilation with a Mapleson C circuit delivered a less accurate and less stable PEEP level compared to a self-inflating bag valve resuscitator. … (more)
- Is Part Of:
- Intensive and critical care nursing. Volume 70(2022)
- Journal:
- Intensive and critical care nursing
- Issue:
- Volume 70(2022)
- Issue Display:
- Volume 70, Issue 2022 (2022)
- Year:
- 2022
- Volume:
- 70
- Issue:
- 2022
- Issue Sort Value:
- 2022-0070-2022-0000
- Page Start:
- Page End:
- Publication Date:
- 2022-06
- Subjects:
- Manual ventilation -- Mapleson C circuit -- PEEP -- Self-inflating bag valve resuscitator
Intensive care nursing -- Periodicals
Critical Illness -- nursing -- Periodicals
Intensive Care -- Periodicals
Nursing Care -- Periodicals
Intensive care nursing
Periodicals
Electronic journals
616.028 - Journal URLs:
- http://www.sciencedirect.com/science/journal/09643397 ↗
http://www.harcourt-international.com/journal ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.iccn.2021.103186 ↗
- Languages:
- English
- ISSNs:
- 0964-3397
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4531.836000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 21334.xml