System for delivering microwave ablation to subcutaneous tumors in small-animals under high-field MRI thermometry guidance. (31st December 2022)
- Record Type:
- Journal Article
- Title:
- System for delivering microwave ablation to subcutaneous tumors in small-animals under high-field MRI thermometry guidance. (31st December 2022)
- Main Title:
- System for delivering microwave ablation to subcutaneous tumors in small-animals under high-field MRI thermometry guidance
- Authors:
- Sebek, Jan
Shrestha, Tej B.
Basel, Matthew T.
Chamani, Faraz
Zeinali, Nooshin
Mali, Ivina
Payne, Macy
Timmerman, Sarah A.
Faridi, Pegah
Pyle, Marla
O'Halloran, Martin
Dennedy, M. Conall
Bossmann, Stefan H.
Prakash, Punit - Abstract:
- Abstract: Purpose: Bio-effects following thermal treatments are a function of the achieved temperature profile in tissue, which can be estimated across tumor volumes with real-time MRI thermometry (MRIT). Here, we report on expansion of a previously developed small-animal microwave hyperthermia system integrated with MRIT for delivering thermal ablation to subcutaneously implanted tumors in mice. Methods: Computational models were employed to assess suitability of the 2.45 GHz microwave applicators for delivering ablation to subcutaneous tumor targets in mice. Phantoms and ex-vivo tissues were heated to temperatures in the range 47–67 °C with custom-made microwave applicators for validating MRIT with the proton resonance frequency shift method against fiberoptic thermometry. HAC15 tumors implanted in nude mice ( n = 6) were ablated in vivo and monitored with MRIT in multiple planes. One day post ablation, animals were euthanized, and excised tumors were processed for viability assessment. Results: Average absolute error between temperatures from fiberoptic sensors and MRIT was 0.6 °C across all ex-vivo ablations. During in-vivo experiments, tumors with volumes ranging between 5.4–35.9 mm 3 (mean 14.2 mm 3 ) were ablated (duration: 103–150 s) to achieve 55 °C at the tumor boundary. Thermal doses ≥240 CEM43 were achieved across 90.7–98.0% of tumor volumes for four cases. Ablations were incomplete for remaining cases, attributed to motion-affected thermometry. ThermalAbstract: Purpose: Bio-effects following thermal treatments are a function of the achieved temperature profile in tissue, which can be estimated across tumor volumes with real-time MRI thermometry (MRIT). Here, we report on expansion of a previously developed small-animal microwave hyperthermia system integrated with MRIT for delivering thermal ablation to subcutaneously implanted tumors in mice. Methods: Computational models were employed to assess suitability of the 2.45 GHz microwave applicators for delivering ablation to subcutaneous tumor targets in mice. Phantoms and ex-vivo tissues were heated to temperatures in the range 47–67 °C with custom-made microwave applicators for validating MRIT with the proton resonance frequency shift method against fiberoptic thermometry. HAC15 tumors implanted in nude mice ( n = 6) were ablated in vivo and monitored with MRIT in multiple planes. One day post ablation, animals were euthanized, and excised tumors were processed for viability assessment. Results: Average absolute error between temperatures from fiberoptic sensors and MRIT was 0.6 °C across all ex-vivo ablations. During in-vivo experiments, tumors with volumes ranging between 5.4–35.9 mm 3 (mean 14.2 mm 3 ) were ablated (duration: 103–150 s) to achieve 55 °C at the tumor boundary. Thermal doses ≥240 CEM43 were achieved across 90.7–98.0% of tumor volumes for four cases. Ablations were incomplete for remaining cases, attributed to motion-affected thermometry. Thermal dose-based ablative tumor coverage agreed with viability assessment of excised tumors. Conclusions: We have developed a system for delivering microwave ablation to subcutaneous tumors in small animals under MRIT guidance and demonstrated its performance in - vivo . … (more)
- Is Part Of:
- International journal of hyperthermia. Volume 39:Number 1(2022)
- Journal:
- International journal of hyperthermia
- Issue:
- Volume 39:Number 1(2022)
- Issue Display:
- Volume 39, Issue 1 (2022)
- Year:
- 2022
- Volume:
- 39
- Issue:
- 1
- Issue Sort Value:
- 2022-0039-0001-0000
- Page Start:
- 584
- Page End:
- 594
- Publication Date:
- 2022-12-31
- Subjects:
- Small-animal ablation system -- magnetic resonance imaging -- microwave ablation -- MRI thermometry
Thermotherapy -- Periodicals
615.832 - Journal URLs:
- http://informahealthcare.com/loi/hth ↗
http://www.tandf.co.uk/journals/titles/02656736.asp ↗
http://informahealthcare.com ↗ - DOI:
- 10.1080/02656736.2022.2061727 ↗
- Languages:
- English
- ISSNs:
- 0265-6736
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.297000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 21289.xml