The effect of tin prefiltration on extremity cone-beam CT imaging with a twin robotic X-ray system. Issue 2 (May 2022)
- Record Type:
- Journal Article
- Title:
- The effect of tin prefiltration on extremity cone-beam CT imaging with a twin robotic X-ray system. Issue 2 (May 2022)
- Main Title:
- The effect of tin prefiltration on extremity cone-beam CT imaging with a twin robotic X-ray system
- Authors:
- Luetkens, K.S.
Huflage, H.
Kunz, A.S.
Ritschl, L.
Herbst, M.
Kappler, S.
Ergün, S.
Goertz, L.
Pennig, L.
Bley, T.A.
Gassenmaier, T.
Grunz, J.-P. - Abstract:
- Abstract: Introduction: While tin prefiltration is established in various CT applications, its value in extremity cone-beam CT relative to optimized spectra has not been thoroughly assessed thus far. This study aims to investigate the effect of tin filters in extremity cone-beam CT with a twin-robotic X-ray system. Methods: Wrist, elbow and ankle joints of two cadaveric specimens were examined in a laboratory setup with different combinations of prefiltration (copper, tin), tube voltage and current–time product. Image quality was assessed subjectively by five radiologists with Fleiss' kappa being computed to measure interrater agreement. To provide a semiquantitative criterion for image quality, contrast-to-noise ratios (CNR) were compared for standardized regions of interest. Volume CT dose indices were calculated for a 16 cm polymethylmethacrylate phantom. Results: Radiation dose ranged from 17.4 mGy in the clinical standard protocol without tin filter to as low as 0.7 mGy with tin prefiltration. Image quality ratings and CNR for tin-filtered scans with 100 kV were lower than for 80 kV studies with copper prefiltration despite higher dose (11.2 and 5.6 vs. 4.5 mGy; p < 0.001). No difference was ascertained between 100 kV scans with tin filtration and 60 kV copper-filtered scans with 75% dose reduction (subjective: p = 0.101; CNR: p = 0.706). Fleiss' kappa of 0.597 (95% confidence interval 0.567–0.626; p < 0.001) indicated moderate interrater agreement. Conclusion:Abstract: Introduction: While tin prefiltration is established in various CT applications, its value in extremity cone-beam CT relative to optimized spectra has not been thoroughly assessed thus far. This study aims to investigate the effect of tin filters in extremity cone-beam CT with a twin-robotic X-ray system. Methods: Wrist, elbow and ankle joints of two cadaveric specimens were examined in a laboratory setup with different combinations of prefiltration (copper, tin), tube voltage and current–time product. Image quality was assessed subjectively by five radiologists with Fleiss' kappa being computed to measure interrater agreement. To provide a semiquantitative criterion for image quality, contrast-to-noise ratios (CNR) were compared for standardized regions of interest. Volume CT dose indices were calculated for a 16 cm polymethylmethacrylate phantom. Results: Radiation dose ranged from 17.4 mGy in the clinical standard protocol without tin filter to as low as 0.7 mGy with tin prefiltration. Image quality ratings and CNR for tin-filtered scans with 100 kV were lower than for 80 kV studies with copper prefiltration despite higher dose (11.2 and 5.6 vs. 4.5 mGy; p < 0.001). No difference was ascertained between 100 kV scans with tin filtration and 60 kV copper-filtered scans with 75% dose reduction (subjective: p = 0.101; CNR: p = 0.706). Fleiss' kappa of 0.597 (95% confidence interval 0.567–0.626; p < 0.001) indicated moderate interrater agreement. Conclusion: Considerable dose reduction is feasible with tin prefiltration, however, the twin-robotic X-ray system's low-dose potential for extremity 3D imaging is maximized with a dedicated low-kilovolt scan protocol in situations without extensive beam-hardening artifacts. Implications for practice: Low-kilovolt imaging with copper prefiltration provides a superior trade-off between dose reduction and image quality compared to tin-filtered cone-beam CT scan protocols with higher tube voltage. … (more)
- Is Part Of:
- Radiography. Volume 28:Issue 2(2022)
- Journal:
- Radiography
- Issue:
- Volume 28:Issue 2(2022)
- Issue Display:
- Volume 28, Issue 2 (2022)
- Year:
- 2022
- Volume:
- 28
- Issue:
- 2
- Issue Sort Value:
- 2022-0028-0002-0000
- Page Start:
- 433
- Page End:
- 439
- Publication Date:
- 2022-05
- Subjects:
- Radiation dosage -- Cone-beam computed tomography -- Extremities -- Tin -- Filtration -- Prefiltration
CBCT Cone-beam computed tomography -- CNR Contrast-to-noise ratio -- CTDIvol(16cm) Volume computed tomography dose index (for 16 cm diameter PMMA dosimetry phantom) -- Cu Scan protocols with standard copper prefiltration -- MDCT Multidetector computed tomography -- ROI Region of interest -- Sn Scan protocols with experimental tin prefiltration
Diagnostic imaging -- Periodicals
Radiotherapy -- Periodicals
Cancer -- Radiotherapy -- Periodicals
Diagnostic Imaging -- Periodicals
Neoplasms -- Periodicals
Radiotherapy -- Periodicals
Imagerie pour le diagnostic -- Périodiques
Radiothérapie -- Périodiques
Cancer -- Radiothérapie -- Périodiques
Electronic journals
616.0757 - Journal URLs:
- http://www.sciencedirect.com/science/journal/10788174 ↗
http://www.radiographyonline.com/ ↗
http://www.harcourt-international.com/journals ↗
http://www.idealibrary.com/links/toc/radi/ ↗
http://www.clinicalkey.com/dura/browse/journalIssue/10788174 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/10788174 ↗
http://www.elsevier.com/journals ↗
http://www.journals.elsevier.com/radiography/ ↗ - DOI:
- 10.1016/j.radi.2021.10.009 ↗
- Languages:
- English
- ISSNs:
- 1078-8174
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7237.001000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 21247.xml