Target margin design through analyzing a large cohort of clinical log data in the cyberknife system. Issue 3 (19th January 2022)
- Record Type:
- Journal Article
- Title:
- Target margin design through analyzing a large cohort of clinical log data in the cyberknife system. Issue 3 (19th January 2022)
- Main Title:
- Target margin design through analyzing a large cohort of clinical log data in the cyberknife system
- Authors:
- Samadi Miandoab, Payam
Saramad, Shahyar
Setayeshi, Saeed - Abstract:
- Abstract: Purpose: Calculating the adequate target margin for real‐time tumor tracking using the Cyberknife system is a challenging issue since different sources of error exist. In this study, the clinical log data of the Cyberknife system were analyzed to adequately quantify the planned target volume (PTV) margins of tumors located in the lung and abdomen regions. Methods: In this study, 45 patients treated with the Cyberknife module were examined. In this context, adequate PTV margins were estimated based on the Van Herk formulation and the uncertainty estimation method by considering the impact of errors and uncertainties. To investigate the impact of errors and uncertainties on the estimated PTV margins, a statistical analysis was also performed. Results: Our study demonstrates five different sources of errors, including segmentation, deformation, correlation, prediction, and targeting errors, which were identified as the main sources of error in the Cyberknife system. Furthermore, the clinical evaluation of the current study reveals that the two different formalisms provided almost identical PTV margin estimates. Additionally, 4–5 mm and 5 mm margins on average could provide adequate PTV margins at lung and abdomen tumors in all three directions, respectively. Overall, it was found that concerning the PTV margins, the impact of correlation and prediction errors is very high, while the impact of robotics error is low. Conclusions: The current study can address twoAbstract: Purpose: Calculating the adequate target margin for real‐time tumor tracking using the Cyberknife system is a challenging issue since different sources of error exist. In this study, the clinical log data of the Cyberknife system were analyzed to adequately quantify the planned target volume (PTV) margins of tumors located in the lung and abdomen regions. Methods: In this study, 45 patients treated with the Cyberknife module were examined. In this context, adequate PTV margins were estimated based on the Van Herk formulation and the uncertainty estimation method by considering the impact of errors and uncertainties. To investigate the impact of errors and uncertainties on the estimated PTV margins, a statistical analysis was also performed. Results: Our study demonstrates five different sources of errors, including segmentation, deformation, correlation, prediction, and targeting errors, which were identified as the main sources of error in the Cyberknife system. Furthermore, the clinical evaluation of the current study reveals that the two different formalisms provided almost identical PTV margin estimates. Additionally, 4–5 mm and 5 mm margins on average could provide adequate PTV margins at lung and abdomen tumors in all three directions, respectively. Overall, it was found that concerning the PTV margins, the impact of correlation and prediction errors is very high, while the impact of robotics error is low. Conclusions: The current study can address two limitations in previous researches, namely insufficient sample sites and a smaller number of patients. A comparison of the present results concerning the lung and abdomen areas with other studies reveals that the proposed strategy could provide a better reference in selection the PTV margins. To our knowledge, this study is one of the first attempts to estimate the PTV margins in the lung and abdomen regions for a large cohort of patients treated using the Cyberknife system. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 23:Issue 3(2022)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 23:Issue 3(2022)
- Issue Display:
- Volume 23, Issue 3 (2022)
- Year:
- 2022
- Volume:
- 23
- Issue:
- 3
- Issue Sort Value:
- 2022-0023-0003-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2022-01-19
- Subjects:
- Cyberknife system -- errors -- PTV margins -- uncertainties
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.13476 ↗