A novel approach to temporary stenting: degradable cardiovascular stents produced from corrodible metal—results 6–18 months after implantation into New Zealand white rabbits. Issue 5 (1st November 2001)
- Record Type:
- Journal Article
- Title:
- A novel approach to temporary stenting: degradable cardiovascular stents produced from corrodible metal—results 6–18 months after implantation into New Zealand white rabbits. Issue 5 (1st November 2001)
- Main Title:
- A novel approach to temporary stenting: degradable cardiovascular stents produced from corrodible metal—results 6–18 months after implantation into New Zealand white rabbits
- Authors:
- Peuster, M
Wohlsein, P
Brügmann, M
Ehlerding, M
Seidler, K
Fink, C
Brauer, H
Fischer, A
Hausdorf, G - Abstract:
- Abstract : OBJECTIVE: To determine whether corrodible materials may be safely used as biodegradable cardiovascular implants. DESIGN: Corrodible iron stents (> 99.8% iron) were produced from pure iron and laser cut with a stent design similar to a commercially available permanent stent (PUVA-AS16). A total of 16 NOR-I stents were implanted into the native descending aorta of 16 New Zealand white rabbits (mean luminal diameter at the implantation site 3.4 mm, balloon diameter to vessel diameter ratio 1.13). RESULTS: No thromboembolic complications and no adverse events occurred during the follow up of 6–18 months. All stents were patent at repeat angiography after 6 (n = 9), 12 (n = 5), and 18 months (n = 2) with no significant neointimal proliferation, no pronounced inflammatory response, and no systemic toxicity. CONCLUSIONS: This initial in vivo experience suggests that degradable iron stents can be safely implanted without significant obstruction of the stented vessel caused by inflammation, neointimal proliferation, or thrombotic events.
- Is Part Of:
- Heart. Volume 86:Issue 5(2001)
- Journal:
- Heart
- Issue:
- Volume 86:Issue 5(2001)
- Issue Display:
- Volume 86, Issue 5 (2001)
- Year:
- 2001
- Volume:
- 86
- Issue:
- 5
- Issue Sort Value:
- 2001-0086-0005-0000
- Page Start:
- 563
- Page End:
- 569
- Publication Date:
- 2001-11-01
- Subjects:
- congenital heart disease -- corrosion -- stents -- biodegradation
Heart -- Diseases -- Treatment -- Periodicals
Cardiology -- Periodicals
616.12 - Journal URLs:
- http://www.bmj.com/archive ↗
http://heart.bmj.com ↗
http://www.heartjnl.com ↗ - DOI:
- 10.1136/heart.86.5.563 ↗
- Languages:
- English
- ISSNs:
- 1355-6037
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 20929.xml