A coordinate positioning puncture method under robot-assisted CT-guidance: phantom and animal experiments. Issue 2 (1st February 2022)
- Record Type:
- Journal Article
- Title:
- A coordinate positioning puncture method under robot-assisted CT-guidance: phantom and animal experiments. Issue 2 (1st February 2022)
- Main Title:
- A coordinate positioning puncture method under robot-assisted CT-guidance: phantom and animal experiments
- Authors:
- Zhang, Weifan
Xia, Peng
Liu, Shijie
Huang, Xiaowei
Zhao, Xinhui
Liu, Zhao
Dang, Hui
Li, Xiaohu
Niu, Gang - Abstract:
- Abstract: Purpose: To evaluate the accuracy of the robot-assisted computed tomography (CT)-guided coordinate positioning puncture method by phantom and animal experiments. Material and methods: In the phantom experiment, seven robot-assisted punctures were made to evaluate the accuracy of the method. In the animal experiment, 18 punctures (nine robotic and nine manual) were made in the livers of nine rabbits. The indicators, such as needle-tract length, angle deviation, puncture accuracy, number of scans required, and radiation exposure dose were compared between manual and robotic punctures. The paired-samples t-test was used for analysis. Results: In the phantom experiment, the mean accuracy of seven punctures was 2.67 mm. In the animal experiment, there was no significant difference in needle-tract length (32.58 mm vs. 34.04 mm, p = .606), angle deviation (17.21° vs. 21.23° p = .557) and puncture accuracy (8.42 vs. 8.77 mm, p = .851) between the two groups. However, the number CT scans required (2.44 vs. 3.33, p = .002), and the radiation exposure dose (772.98 vs. 1077.89 mGy/cm, p = .003) were lower in the robot group than in the manual group. Conclusions: The coordinate positioning puncture method under robot-assisted CT-guidance can reach an accuracy that is comparable to that of the traditional manual CT-guided puncture method and with fewer CT scanning times accompanied with a lower radiation dosage.
- Is Part Of:
- Minimally invasive therapy & allied technologies. Volume 31:Issue 2(2022)
- Journal:
- Minimally invasive therapy & allied technologies
- Issue:
- Volume 31:Issue 2(2022)
- Issue Display:
- Volume 31, Issue 2 (2022)
- Year:
- 2022
- Volume:
- 31
- Issue:
- 2
- Issue Sort Value:
- 2022-0031-0002-0000
- Page Start:
- 206
- Page End:
- 215
- Publication Date:
- 2022-02-01
- Subjects:
- Robot-assisted -- CT-guidance -- puncture -- coordinate positioning -- animal experiment
Endoscopy -- Periodicals
Interventional radiology -- Periodicals
Endoscopic surgery -- Periodicals
617.05 - Journal URLs:
- http://informahealthcare.com/loi/mit ↗
http://informahealthcare.com ↗ - DOI:
- 10.1080/13645706.2020.1787451 ↗
- Languages:
- English
- ISSNs:
- 1364-5706
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5797.714000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 20789.xml