Image Quality Assessment for Clinical Cadmium Telluride-Based Photon-Counting Computed Tomography Detector in Cadaveric Wrist Imaging. Issue 12 (December 2021)
- Record Type:
- Journal Article
- Title:
- Image Quality Assessment for Clinical Cadmium Telluride-Based Photon-Counting Computed Tomography Detector in Cadaveric Wrist Imaging. Issue 12 (December 2021)
- Main Title:
- Image Quality Assessment for Clinical Cadmium Telluride-Based Photon-Counting Computed Tomography Detector in Cadaveric Wrist Imaging
- Authors:
- Grunz, Jan-Peter
Huflage, Henner
Heidenreich, Julius Frederik
Ergün, Süleyman
Petersilka, Martin
Allmendinger, Thomas
Bley, Thorsten Alexander
Petritsch, Bernhard - Abstract:
- Abstract : Objectives: Detailed visualization of bone microarchitecture is essential for assessment of wrist fractures in computed tomography (CT). This study aims to evaluate the imaging performance of a CT system with clinical cadmium telluride-based photon-counting detector (PCD-CT) compared with a third-generation dual-source CT scanner with energy-integrating detector technology (EID-CT). Materials and Methods: Both CT systems were used for the examination of 8 cadaveric wrists with radiation dose equivalent scan protocols (low-/standard-/full-dose imaging: CTDIvol = 1.50/5.80/8.67 mGy). All wrists were scanned with 2 different operating modes of the photon-counting CT (standard-resolution and ultra-high-resolution). After reformatting with comparable reconstruction parameters and convolution kernels, subjective evaluation of image quality was performed by 3 radiologists on a 7-point scale. For estimation of interrater reliability, we report the intraclass correlation coefficient (absolute agreement, 2-way random-effects model). Signal-to-noise and contrast-to-noise ratios were calculated to provide semiquantitative assessment of image quality. Results: Subjective image quality of standard-dose PCD-CT examinations in ultra-high-resolution mode was superior compared with full-dose PCD-CT in standard-resolution mode ( P = 0.016) and full-dose EID-CT ( P = 0.040). No difference was ascertained between low-dose PCD-CT in ultra-high-resolution mode and standard-dose scansAbstract : Objectives: Detailed visualization of bone microarchitecture is essential for assessment of wrist fractures in computed tomography (CT). This study aims to evaluate the imaging performance of a CT system with clinical cadmium telluride-based photon-counting detector (PCD-CT) compared with a third-generation dual-source CT scanner with energy-integrating detector technology (EID-CT). Materials and Methods: Both CT systems were used for the examination of 8 cadaveric wrists with radiation dose equivalent scan protocols (low-/standard-/full-dose imaging: CTDIvol = 1.50/5.80/8.67 mGy). All wrists were scanned with 2 different operating modes of the photon-counting CT (standard-resolution and ultra-high-resolution). After reformatting with comparable reconstruction parameters and convolution kernels, subjective evaluation of image quality was performed by 3 radiologists on a 7-point scale. For estimation of interrater reliability, we report the intraclass correlation coefficient (absolute agreement, 2-way random-effects model). Signal-to-noise and contrast-to-noise ratios were calculated to provide semiquantitative assessment of image quality. Results: Subjective image quality of standard-dose PCD-CT examinations in ultra-high-resolution mode was superior compared with full-dose PCD-CT in standard-resolution mode ( P = 0.016) and full-dose EID-CT ( P = 0.040). No difference was ascertained between low-dose PCD-CT in ultra-high-resolution mode and standard-dose scans with either PCD-CT in standard-resolution mode ( P = 0.108) or EID-CT ( P = 0.470). Observer evaluation of standard-resolution PCD-CT and EID-CT delivered similar results in full- and standard-dose scans ( P = 0.248/0.509). Intraclass correlation coefficient was 0.876 (95% confidence interval, 0.744–0.925; P < 0.001), indicating good reliability. Between dose equivalent studies, signal-to-noise and contrast-to-noise ratios were substantially higher in photon-counting CT examinations (all P 's < 0.001). Conclusions: Superior visualization of fine anatomy is feasible with the clinical photon-counting CT system in cadaveric wrist scans. The ultra-high-resolution scan mode suggests potential for considerable dose reduction over energy-integrating dual-source CT. … (more)
- Is Part Of:
- Investigative radiology. Volume 56:Issue 12(2021)
- Journal:
- Investigative radiology
- Issue:
- Volume 56:Issue 12(2021)
- Issue Display:
- Volume 56, Issue 12 (2021)
- Year:
- 2021
- Volume:
- 56
- Issue:
- 12
- Issue Sort Value:
- 2021-0056-0012-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-12
- Subjects:
- photon-counting -- tomography -- x-ray computed -- wrist -- radiation dosage -- cancellous bone
Diagnosis, Radioscopic -- Periodicals
Radiology, Medical -- Periodicals
616.0757 - Journal URLs:
- http://journals.lww.com/investigativeradiology/pages/default.aspx ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/RLI.0000000000000789 ↗
- Languages:
- English
- ISSNs:
- 0020-9996
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4560.350000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 20682.xml