A comparison of relative humidity calibration facilities at temperatures up to 170 °C. (15th February 2022)
- Record Type:
- Journal Article
- Title:
- A comparison of relative humidity calibration facilities at temperatures up to 170 °C. (15th February 2022)
- Main Title:
- A comparison of relative humidity calibration facilities at temperatures up to 170 °C
- Authors:
- Peruzzi, A.
Bosma, R.
Tabandeh, S.
Fernicola, V.
Georgin, E. - Abstract:
- Highlights: SI-traceability of relative humidity extended to very high dew-point temperatures. Degree of equivalence of three national humidity measurement standards. Demonstrated relative humidity calibrations at very high dew-point temperatures. Abstract: Three European National Metrology Institutes (NMIs), VSL, LNE-CETIAT and INRiM, extended their relative humidity calibration capabilities to temperatures up to 170 °C, dew-point temperatures up to 150 °C and pressures up to 600 kPa. In order to test the equivalence of the respective calibration set-ups, 4 commercial high-temperature relative humidity sensors were exchanged among the three NMIs and calibrated at different combinations of relative humidity (19 %rh to 70 %rh), air temperature (100 °C to 170 °C) and pressure (100 kPa to 500 kPa), realizing dew-point temperatures of either 81.5 °C or 127 °C. The initial and final measurements at VSL showed that the stability of the selected sensors was sufficient for comparing the different calibration set-ups. The results of the comparison showed that VSL and LNE-CETIAT facilities agreed within their claimed uncertainties (≤0.3 %rh), while discrepancies of up to 4 %rh were observed between VSL and INRiM facilities.
- Is Part Of:
- Measurement. Volume 189(2022)
- Journal:
- Measurement
- Issue:
- Volume 189(2022)
- Issue Display:
- Volume 189, Issue 2022 (2022)
- Year:
- 2022
- Volume:
- 189
- Issue:
- 2022
- Issue Sort Value:
- 2022-0189-2022-0000
- Page Start:
- Page End:
- Publication Date:
- 2022-02-15
- Subjects:
- Relative humidity up to 170 °C -- Sensors for relative humidity up to 170 °C -- Dew-point temperature above 100 °C
Weights and measures -- Periodicals
Measurement -- Periodicals
Measurement
Weights and measures
Periodicals
530.8 - Journal URLs:
- http://www.sciencedirect.com/science/journal/02632241 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.measurement.2021.110435 ↗
- Languages:
- English
- ISSNs:
- 0263-2241
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5413.544700
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 20636.xml