Cardiac output estimation by pulse wave analysis using the pressure recording analytical method and intermittent pulmonary artery thermodilution: A method comparison study after off-pump coronary artery bypass surgery. Issue 10 (October 2020)
- Record Type:
- Journal Article
- Title:
- Cardiac output estimation by pulse wave analysis using the pressure recording analytical method and intermittent pulmonary artery thermodilution: A method comparison study after off-pump coronary artery bypass surgery. Issue 10 (October 2020)
- Main Title:
- Cardiac output estimation by pulse wave analysis using the pressure recording analytical method and intermittent pulmonary artery thermodilution
- Authors:
- Greiwe, Gillis
Luehsen, Katharina
Hapfelmeier, Alexander
Rogge, Dorothea
Kubik, Mathias
Schulte-Uentrop, Leonie
Saugel, Bernd - Abstract:
- Abstract : BACKGROUND: Invasive pulse wave analysis is used in peri-operative settings to estimate cardiac output (CO). The 'pressure recording analytical method' (PRAM) implemented in the MostCareUp CO monitor is an invasive pulse wave analysis method using high-frequency sampling and analysis of the pulse wave to directly estimate the arterial impedance as a key variable of the proprietary CO estimation algorithm. OBJECTIVE: To compare CO estimated by PRAM (PRAM-CO; test method) with CO measured by pulmonary artery thermodilution (PATD-CO; reference method). DESIGN: Prospective observational method comparison study. PRAM-CO and PATD-CO were assessed simultaneously at five time points with at least 20 min between measurements. Arterial pressure waveforms were carefully checked for damping artefacts and a proprietary electronic filter of the MostCareUp CO monitor was used to optimise waveform quality. SETTING: ICU of a German university hospital from August 2018 until April 2019. PATIENTS: We included adult patients admitted to the ICU after elective off-pump coronary artery bypass surgery who were monitored with a radial arterial catheter and a pulmonary artery catheter. Patients with severe heart valve insufficiency or persistent arrhythmia were excluded. MAIN OUTCOME MEASURES AND ANALYSIS: PATD-CO and PRAM-CO were compared using Bland-Altman analysis accounting for repeated measurements, the percentage error and trending analysis (four-quadrant plot, concordance rate).Abstract : BACKGROUND: Invasive pulse wave analysis is used in peri-operative settings to estimate cardiac output (CO). The 'pressure recording analytical method' (PRAM) implemented in the MostCareUp CO monitor is an invasive pulse wave analysis method using high-frequency sampling and analysis of the pulse wave to directly estimate the arterial impedance as a key variable of the proprietary CO estimation algorithm. OBJECTIVE: To compare CO estimated by PRAM (PRAM-CO; test method) with CO measured by pulmonary artery thermodilution (PATD-CO; reference method). DESIGN: Prospective observational method comparison study. PRAM-CO and PATD-CO were assessed simultaneously at five time points with at least 20 min between measurements. Arterial pressure waveforms were carefully checked for damping artefacts and a proprietary electronic filter of the MostCareUp CO monitor was used to optimise waveform quality. SETTING: ICU of a German university hospital from August 2018 until April 2019. PATIENTS: We included adult patients admitted to the ICU after elective off-pump coronary artery bypass surgery who were monitored with a radial arterial catheter and a pulmonary artery catheter. Patients with severe heart valve insufficiency or persistent arrhythmia were excluded. MAIN OUTCOME MEASURES AND ANALYSIS: PATD-CO and PRAM-CO were compared using Bland-Altman analysis accounting for repeated measurements, the percentage error and trending analysis (four-quadrant plot, concordance rate). RESULTS: We analysed 195 paired CO values of 41 patients. Mean PATD-CO and PRAM-CO were 4.99 ± 1.02 and 4.92 ± 1.05 l min −1, respectively. PATD-CO and PRAM-CO ranged from 3.04 to 8.74 and 2.79 to 8.01 l min −1, respectively. The mean of the differences between PATD-CO and PRAM-CO was −0.08 ± 0.74 l min −1 with 95% limits of agreement of −1.55 to +1.40 l min −1 . The percentage error was 29.8%. The concordance rate in four-quadrant plot analysis was 92%. CONCLUSION: Using the system's electronic waveform filter PRAM-CO shows good agreement and trending ability compared with PATD-CO in adults after off-pump coronary artery bypass surgery. … (more)
- Is Part Of:
- European journal of anaesthesiology. Volume 37:Issue 10(2020:Oct.)
- Journal:
- European journal of anaesthesiology
- Issue:
- Volume 37:Issue 10(2020:Oct.)
- Issue Display:
- Volume 37, Issue 10 (2020)
- Year:
- 2020
- Volume:
- 37
- Issue:
- 10
- Issue Sort Value:
- 2020-0037-0010-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-10
- Subjects:
- Anesthesiology -- Periodicals
Anesthesiology -- Periodicals
Anesthésiologie -- Périodiques
Anesthesiology
Periodicals
Electronic journals
617.96 - Journal URLs:
- http://journals.lww.com/ejanaesthesiology/pages/default.aspx ↗
http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1365-2346/issues ↗
http://www.blackwell-synergy.com/member/institutions/issuelist.asp?journal=eja ↗
http://ovidsp.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&PAGE=toc&D=ovft&AN=00003643-000000000-00000 ↗
http://journals.lww.com ↗
http://www.lww.com/Product/0265-0215 ↗ - DOI:
- 10.1097/EJA.0000000000001227 ↗
- Languages:
- English
- ISSNs:
- 0265-0215
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.722200
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 20525.xml