Carotid Doppler Measurement Variability in Functional Hemodynamic Monitoring: An Analysis of 17, 822 Cardiac Cycles. (11th June 2021)
- Record Type:
- Journal Article
- Title:
- Carotid Doppler Measurement Variability in Functional Hemodynamic Monitoring: An Analysis of 17, 822 Cardiac Cycles. (11th June 2021)
- Main Title:
- Carotid Doppler Measurement Variability in Functional Hemodynamic Monitoring: An Analysis of 17, 822 Cardiac Cycles
- Authors:
- Kenny, Jon-Émile S.
Barjaktarevic, Igor
Mackenzie, David C.
Elfarnawany, Mai
Math, Zhen Yang B.
Eibl, Andrew M.
Eibl, Joseph K.
Kim, Chul Ho
Johnson, Bruce D. - Abstract:
- Abstract : OBJECTIVES: Carotid Doppler ultrasound is used as a measure of fluid responsiveness, however, assessing change with statistical confidence requires an adequate beat sample size. The coefficient of variation helps quantify the number of cardiac cycles needed to adequately detect change during functional hemodynamic monitoring. DESIGN: Prospective, observational, human model of hemorrhage and resuscitation. SETTING: Human physiology laboratory at Mayo Clinic. SUBJECTS: Healthy volunteers. INTERVENTIONS: Lower body negative pressure. MEASUREMENTS AND MAIN RESULTS: We measured the coefficient of variation of the carotid artery velocity time integral and corrected flow time during significant cardiac preload changes. Seventeen-thousand eight-hundred twenty-two cardiac cycles were analyzed. The median coefficient of variation of the carotid velocity time integral was 8.7% at baseline and 11.9% during lowest-tolerated lower body negative pressure stage. These values were 3.6% and 4.6%, respectively, for the corrected flow time. CONCLUSIONS: The median coefficient of variation values measured in this large dataset indicates that at least 6 cardiac cycles should be averaged before and after an intervention when using the carotid artery as a functional hemodynamic measure.
- Is Part Of:
- Critical care explorations. Volume 3:Number 6(2021)
- Journal:
- Critical care explorations
- Issue:
- Volume 3:Number 6(2021)
- Issue Display:
- Volume 3, Issue 6 (2021)
- Year:
- 2021
- Volume:
- 3
- Issue:
- 6
- Issue Sort Value:
- 2021-0003-0006-0000
- Page Start:
- e0439
- Page End:
- Publication Date:
- 2021-06-11
- Subjects:
- carotid Doppler -- corrected flow time -- fluid responsiveness -- measurement variability -- velocity time integral
- Journal URLs:
- http://journals.lww.com/pages/default.aspx ↗
- DOI:
- 10.1097/CCE.0000000000000439 ↗
- Languages:
- English
- ISSNs:
- 2639-8028
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 20446.xml