Development of a micro-scale method to assess the effect of corrosion on the mechanical properties of a biodegradable Fe-316L stent material. (February 2021)
- Record Type:
- Journal Article
- Title:
- Development of a micro-scale method to assess the effect of corrosion on the mechanical properties of a biodegradable Fe-316L stent material. (February 2021)
- Main Title:
- Development of a micro-scale method to assess the effect of corrosion on the mechanical properties of a biodegradable Fe-316L stent material
- Authors:
- Frattolin, Jennifer
Cattarinuzzi, Emanuele
Rajagopalan, Sriraman
Gastaldi, Dario
Vena, Pasquale
Yue, Stephen
Bertrand, Olivier F.
Mongrain, Rosaire - Abstract:
- Abstract: The application of biodegradable materials to stent design has the potential to transform coronary artery disease treatment. It is critical that biodegradable stents have sustained strength during degradation and vessel healing to prevent re-occlusion. Proper assessment of the impact of corrosion on the mechanical behaviour of potential biomaterials is important. Investigations within literature frequently implement simplified testing conditions to understand this behaviour and fail to consider size effects associated with strut thickness, or the increase in corrosion due to blood flow, both of which can impact material properties. A protocol was developed that utilizes micro-scale specimens, in conjunction with dynamic degradation, to assess the effect of corrosion on the mechanical properties of a novel Fe-316L material. Dynamic degradation led to increased specimen corrosion, resulting in a greater reduction in strength after 48 h of degradation in comparison to samples statically corroded. It was found that thicker micro-tensile samples (h > 200 μm) had a greater loss of strength in comparison to its thinner counterpart (h < 200 μm), due to increased corrosion of the thicker samples (203 MPa versus 260 MPa after 48 h, p = 0.0017). This investigation emphasizes the necessity of implementing physiologically relevant testing conditions, including dynamic corrosion and stent strut thickness, when evaluating potential biomaterials for biodegradable stentAbstract: The application of biodegradable materials to stent design has the potential to transform coronary artery disease treatment. It is critical that biodegradable stents have sustained strength during degradation and vessel healing to prevent re-occlusion. Proper assessment of the impact of corrosion on the mechanical behaviour of potential biomaterials is important. Investigations within literature frequently implement simplified testing conditions to understand this behaviour and fail to consider size effects associated with strut thickness, or the increase in corrosion due to blood flow, both of which can impact material properties. A protocol was developed that utilizes micro-scale specimens, in conjunction with dynamic degradation, to assess the effect of corrosion on the mechanical properties of a novel Fe-316L material. Dynamic degradation led to increased specimen corrosion, resulting in a greater reduction in strength after 48 h of degradation in comparison to samples statically corroded. It was found that thicker micro-tensile samples (h > 200 μm) had a greater loss of strength in comparison to its thinner counterpart (h < 200 μm), due to increased corrosion of the thicker samples (203 MPa versus 260 MPa after 48 h, p = 0.0017). This investigation emphasizes the necessity of implementing physiologically relevant testing conditions, including dynamic corrosion and stent strut thickness, when evaluating potential biomaterials for biodegradable stent application. Graphical abstract: Image 1 Highlights: Development of micro-tensile protocol in conjunction with dynamic corrosion tests. Effect of corrosion on mechanical properties of Fe-316L influenced by specimen size. Dynamic degradation had greater impact than static corrosion environment. Thicker micro-samples had greater reduction in strength than thinner micro-samples. … (more)
- Is Part Of:
- Journal of the mechanical behavior of biomedical materials. Volume 114(2021)
- Journal:
- Journal of the mechanical behavior of biomedical materials
- Issue:
- Volume 114(2021)
- Issue Display:
- Volume 114, Issue 2021 (2021)
- Year:
- 2021
- Volume:
- 114
- Issue:
- 2021
- Issue Sort Value:
- 2021-0114-2021-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-02
- Subjects:
- Biodegradable stent -- Corrosion -- Mechanical properties -- Degradation -- Micro-tensile -- Cold-gas dynamic spraying
Biomedical materials -- Periodicals
Biomedical materials -- Mechanical properties -- Periodicals
Biomedical materials
Biomedical materials -- Mechanical properties
Periodicals
Electronic journals
610.28 - Journal URLs:
- http://www.sciencedirect.com/science/journal/17516161 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jmbbm.2020.104173 ↗
- Languages:
- English
- ISSNs:
- 1751-6161
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5015.809000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 20405.xml