The choice of the femoral center of rotation affects material loss in total knee replacement wear testing – A parametric finite element study of ISO 14243-3. (9th May 2019)
- Record Type:
- Journal Article
- Title:
- The choice of the femoral center of rotation affects material loss in total knee replacement wear testing – A parametric finite element study of ISO 14243-3. (9th May 2019)
- Main Title:
- The choice of the femoral center of rotation affects material loss in total knee replacement wear testing – A parametric finite element study of ISO 14243-3
- Authors:
- Mell, Steven P.
Wimmer, Markus A.
Lundberg, Hannah J. - Abstract:
- Abstract: A leading cause of long-term failure of total knee replacements (TKRs) is osteolysis caused by polyethylene wear particles. The current gold standard for preclinical wear testing of TKRs is mechanical knee simulators. The definition of the femoral center of flexion-extension rotation (CoR) has been identified as one possible source of variability within TKR wear tests, since the femoral curvature varies from distal to posterior. The magnitude of the influence on wear due to changes in location of femoral CoR has not been investigated in depth. During this study, a computational framework utilizing finite element analysis for modelling wear of TKRs was developed and used to investigate the influence of the location of femoral CoR on TKR polyethylene wear during standardized displacement controlled testing (ISO 14243-3:2014). The study was carried out using a 40-point Latin Hypercube Design of Experiments approach. Volumetric wear was highly correlated to femoral CoR in both the superior/inferior and anterior/posterior directions, with a stronger relationship in the superior/inferior direction. In addition, wear scars showing linear penetration were examined, with large differences in simulations at the extreme ends of the sampling region. In this study, it was found that variations in the location of the femoral center of rotation can represent a large source of variability in the preclinical testing and evaluation of the wear performance of total knee replacements.Abstract: A leading cause of long-term failure of total knee replacements (TKRs) is osteolysis caused by polyethylene wear particles. The current gold standard for preclinical wear testing of TKRs is mechanical knee simulators. The definition of the femoral center of flexion-extension rotation (CoR) has been identified as one possible source of variability within TKR wear tests, since the femoral curvature varies from distal to posterior. The magnitude of the influence on wear due to changes in location of femoral CoR has not been investigated in depth. During this study, a computational framework utilizing finite element analysis for modelling wear of TKRs was developed and used to investigate the influence of the location of femoral CoR on TKR polyethylene wear during standardized displacement controlled testing (ISO 14243-3:2014). The study was carried out using a 40-point Latin Hypercube Design of Experiments approach. Volumetric wear was highly correlated to femoral CoR in both the superior/inferior and anterior/posterior directions, with a stronger relationship in the superior/inferior direction. In addition, wear scars showing linear penetration were examined, with large differences in simulations at the extreme ends of the sampling region. In this study, it was found that variations in the location of the femoral center of rotation can represent a large source of variability in the preclinical testing and evaluation of the wear performance of total knee replacements. This study represents the first attempt at quantifying the effect on wear of different femoral center of rotations across a large sampling space. … (more)
- Is Part Of:
- Journal of biomechanics. Volume 88(2019)
- Journal:
- Journal of biomechanics
- Issue:
- Volume 88(2019)
- Issue Display:
- Volume 88, Issue 2019 (2019)
- Year:
- 2019
- Volume:
- 88
- Issue:
- 2019
- Issue Sort Value:
- 2019-0088-2019-0000
- Page Start:
- 104
- Page End:
- 112
- Publication Date:
- 2019-05-09
- Subjects:
- Total knee arthroplasty -- Computational modelling -- Polyethylene wear -- Finite element analysis
Animal mechanics -- Periodicals
Biomechanics -- Periodicals
Biomechanics -- Periodicals
Mécanique animale -- Périodiques
Biomécanique -- Périodiques
Electronic journals
571.4305 - Journal URLs:
- http://www.sciencedirect.com/science/journal/00219290 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/00219290 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/00219290 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jbiomech.2019.03.027 ↗
- Languages:
- English
- ISSNs:
- 0021-9290
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4953.600000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 20386.xml