Prophylactic augmentation implants in the proximal femur for hip fracture prevention: An in silico investigation of simulated sideways fall impacts. (February 2022)
- Record Type:
- Journal Article
- Title:
- Prophylactic augmentation implants in the proximal femur for hip fracture prevention: An in silico investigation of simulated sideways fall impacts. (February 2022)
- Main Title:
- Prophylactic augmentation implants in the proximal femur for hip fracture prevention: An in silico investigation of simulated sideways fall impacts
- Authors:
- Fung, Anita
Fleps, Ingmar
Cripton, Peter A.
Guy, Pierre
Ferguson, Stephen J.
Helgason, Benedikt - Abstract:
- Abstract: Femoral fractures from sideways falls in the elderly are associated with significant rates of morbidity and mortality. Approaches to prevent these catastrophic injuries include pharmacological treatments, which have limited efficacy. Prophylactic femoral augmentation systems are a promising alternative that are gaining prominence by addressing the most debilitating osteoporosis-related fracture. We have developed finite element models (FEMs) of a novel experimental sideways fall simulator for cadavers. By virtue of the range of specimens and injury outcomes, these FEMs are well-suited to the evaluation of such implants. The purpose of this study was to use the FEMs to evaluate the mechanical effectiveness of three different prophylactic femoral augmentation systems. Models of the Y-Strut® (Hyprevention®, Pessac, France), Gamma Nail® (Stryker, Kalamazoo, USA), and a simple lag screw femoral fracture implant systems were placed into FEMs of five cadaveric pelvis-femur constructs embedded in a soft tissue surrogate, which were then subject to simulated sideways falls at seven impact velocities. Femur-only FEMs were also evaluated. Peak impact forces and peak acetabular forces were examined, and failure was evaluated using a strain-based criterion. We found that the femoral augmentation systems increased the peak forces prior to fracture, but were unable to prevent fracture for severe impacts. The Gamma Nail® system consistently produced the largest strength increasesAbstract: Femoral fractures from sideways falls in the elderly are associated with significant rates of morbidity and mortality. Approaches to prevent these catastrophic injuries include pharmacological treatments, which have limited efficacy. Prophylactic femoral augmentation systems are a promising alternative that are gaining prominence by addressing the most debilitating osteoporosis-related fracture. We have developed finite element models (FEMs) of a novel experimental sideways fall simulator for cadavers. By virtue of the range of specimens and injury outcomes, these FEMs are well-suited to the evaluation of such implants. The purpose of this study was to use the FEMs to evaluate the mechanical effectiveness of three different prophylactic femoral augmentation systems. Models of the Y-Strut® (Hyprevention®, Pessac, France), Gamma Nail® (Stryker, Kalamazoo, USA), and a simple lag screw femoral fracture implant systems were placed into FEMs of five cadaveric pelvis-femur constructs embedded in a soft tissue surrogate, which were then subject to simulated sideways falls at seven impact velocities. Femur-only FEMs were also evaluated. Peak impact forces and peak acetabular forces were examined, and failure was evaluated using a strain-based criterion. We found that the femoral augmentation systems increased the peak forces prior to fracture, but were unable to prevent fracture for severe impacts. The Gamma Nail® system consistently produced the largest strength increases relative to the unaugmented femur for all five specimens in both the pendulum-drop FEMs and the femur-only simulations. In some cases, the same implant appeared to cause fractures in the acetabulum. The femur-only FEMs showed larger force increases than the pendulum-drop simulations, which suggests that the results of the femur-only simulations may not represent sideways falls as accurately as the soft tissue-embedded pendulum-drop simulations. The results from this study demonstrate the ability to simulate a high energy phenomenon and the effect of implants in an in silico environment. The results also suggest that implants could increase the force applied to the proximal femur during impact. Fracture outcomes from the tested implants can be used to inform the design of future devices, which reaffirms the value of modelling with biofidelic considerations in the implant design process. To the authors' knowledge, this is the first paper to use more complex biofidelic FEMs to assess prophylactic femoral augmentation methods. Graphical abstract: Image 1 Highlights: First study to assess femoral augmentation implants using a complex biofidelic model. Femoral augmentation implants increase the force prior to fracture. Gamma Nail® implant produced the highest strength increases. All specimens fractured at an impact velocity of 2.5 m/s and higher. Strengthening of the femur may cause fractures at the acetabulum. … (more)
- Is Part Of:
- Journal of the mechanical behavior of biomedical materials. Volume 126(2022)
- Journal:
- Journal of the mechanical behavior of biomedical materials
- Issue:
- Volume 126(2022)
- Issue Display:
- Volume 126, Issue 2022 (2022)
- Year:
- 2022
- Volume:
- 126
- Issue:
- 2022
- Issue Sort Value:
- 2022-0126-2022-0000
- Page Start:
- Page End:
- Publication Date:
- 2022-02
- Subjects:
- Mechanical properties -- Femoral implant -- Computational models -- Impact
Biomedical materials -- Periodicals
Biomedical materials -- Mechanical properties -- Periodicals
Biomedical materials
Biomedical materials -- Mechanical properties
Periodicals
Electronic journals
610.28 - Journal URLs:
- http://www.sciencedirect.com/science/journal/17516161 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jmbbm.2021.104957 ↗
- Languages:
- English
- ISSNs:
- 1751-6161
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5015.809000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 20350.xml