Early unexpected failure of a vitamin E-infused highly cross-linked polyethylene liner: A case report. Issue 41 (15th October 2021)
- Record Type:
- Journal Article
- Title:
- Early unexpected failure of a vitamin E-infused highly cross-linked polyethylene liner: A case report. Issue 41 (15th October 2021)
- Main Title:
- Early unexpected failure of a vitamin E-infused highly cross-linked polyethylene liner
- Authors:
- Kim, Kyeong Baek
Lee, Sang-Min
Moon, Nam Hoon
Do, Min Uk
Shin, Won Chul - Other Names:
- Saranathan. Maya section editor.
- Abstract:
- Abstract: Rationale: Total hip arthroplasty (THA) with a polyethylene (PE) liner is 1 of the most effective and successful treatment strategies for end-stage hip disorders. Vitamin E-infused highly cross-linked polyethylene is theoretically known to prevent failure due to oxidative degradation in the body, and is resistant to wear; therefore, successful long-term survival of THA is expected. Patient concerns: In June 2019, approximately 1.5 years after THA, the patient sat down and stood up without any specific issue; however, an unusual bullet sound occurred around the left hip joint. Since then, his discomfort persisted, and he was admitted to the emergency department. Diagnosis: Plain radiography and metal artifact reduction computed tomography performed in the emergency department revealed eccentric elevation of the prosthetic femoral head and suspected PE liner failure. Intervention: Revision surgery was performed for modular component exchange. To reduce the dislocation risk without performing cup exchange, conversion to dual-mobility articulation was performed. Outcomes: During the regular follow-up post-surgery, the patient could perform daily life activities without any discomfort, and dislocation was not observed. At 2 years postoperatively, no significant changes were observed in the radiographic images. Lessons: This case report presents an unexpected failure of THA due to superior rim fracture and excessive wear at the locking mechanism of the vitamin E-infusedAbstract: Rationale: Total hip arthroplasty (THA) with a polyethylene (PE) liner is 1 of the most effective and successful treatment strategies for end-stage hip disorders. Vitamin E-infused highly cross-linked polyethylene is theoretically known to prevent failure due to oxidative degradation in the body, and is resistant to wear; therefore, successful long-term survival of THA is expected. Patient concerns: In June 2019, approximately 1.5 years after THA, the patient sat down and stood up without any specific issue; however, an unusual bullet sound occurred around the left hip joint. Since then, his discomfort persisted, and he was admitted to the emergency department. Diagnosis: Plain radiography and metal artifact reduction computed tomography performed in the emergency department revealed eccentric elevation of the prosthetic femoral head and suspected PE liner failure. Intervention: Revision surgery was performed for modular component exchange. To reduce the dislocation risk without performing cup exchange, conversion to dual-mobility articulation was performed. Outcomes: During the regular follow-up post-surgery, the patient could perform daily life activities without any discomfort, and dislocation was not observed. At 2 years postoperatively, no significant changes were observed in the radiographic images. Lessons: This case report presents an unexpected failure of THA due to superior rim fracture and excessive wear at the locking mechanism of the vitamin E-infused highly cross-linked polyethylene liner. This is an interesting case, as early PE liner failure occurred without strong labor intensity or trauma. A modular component exchange was performed with a dual-mobility bearing, and no issues were observed approximately 2 years after the reoperation. Therefore, third-generation highly cross-linked polyethylene liners can also cause early failure without a clear cause, and this case report highlights the necessity of considering several strategies for reoperation. Study Design: Case report. … (more)
- Is Part Of:
- Medicine. Volume 100:Issue 41(2021)
- Journal:
- Medicine
- Issue:
- Volume 100:Issue 41(2021)
- Issue Display:
- Volume 100, Issue 41 (2021)
- Year:
- 2021
- Volume:
- 100
- Issue:
- 41
- Issue Sort Value:
- 2021-0100-0041-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-10-15
- Subjects:
- dual-mobility -- liner -- polyethylene -- rim fracture -- vitamin E-infused -- wear
Medicine -- Periodicals
Medicine -- Periodicals
Médecine -- Périodiques
Geneeskunde
Medicine
Periodicals
Periodicals
610.5 - Journal URLs:
- http://journals.lww.com/md-journal/pages/default.aspx ↗
http://gateway.ovid.com/ovidweb.cgi?T=JS&PAGE=toc&D=ovft&MODE=ovid&NEWS=N&AN=00002060-000000000-00000 ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/MD.0000000000027454 ↗
- Languages:
- English
- ISSNs:
- 0025-7974
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5534.000000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 20288.xml