End-to-end test for fractionated online adaptive MR-guided radiotherapy using a deformable anthropomorphic pelvis phantom. (16th December 2021)
- Record Type:
- Journal Article
- Title:
- End-to-end test for fractionated online adaptive MR-guided radiotherapy using a deformable anthropomorphic pelvis phantom. (16th December 2021)
- Main Title:
- End-to-end test for fractionated online adaptive MR-guided radiotherapy using a deformable anthropomorphic pelvis phantom
- Authors:
- Elter, A
Rippke, C
Johnen, W
Mann, P
Hellwich, E
Schwahofer, A
Dorsch, S
Buchele, C
Klüter, S
Karger, C P - Abstract:
- Abstract: Objective. In MR-guided radiotherapy (MRgRT) for prostate cancer treatments inter-fractional anatomy changes such as bladder and rectum fillings may be corrected by an online adaption of the treatment plan. To clinically implement such complex treatment procedures, however, specific end-to-end tests are required that are able to validate the overall accuracy of all treatment steps from pre-treatment imaging to dose delivery. Approach. In this study, an end-to-end test of a fractionated and online adapted MRgRT prostate irradiation was performed using the so-called ADAM-PETer phantom. The phantom was adapted to perform 3D polymer gel (PG) dosimetry in the prostate and rectum. Furthermore, thermoluminescence detectors (TLDs) were placed at the center and on the surface of the prostate for additional dose measurements as well as for an external dose renormalization of the PG. For the end-to-end test, a total of five online adapted irradiations were applied in sequence with different bladder and rectum fillings, respectively. Main results. A good agreement of measured and planned dose was found represented by high γ -index passing rates ( 3 % / 3 mm criterion) of the PG evaluation of 98.9 % in the prostate and 93.7 % in the rectum. TLDs used for PG renormalization at the center of the prostate showed a deviation of − 2.3 % . Significance. The presented end-to-end test, which allows for 3D dose verification in the prostate and rectum, demonstrates the feasibility andAbstract: Objective. In MR-guided radiotherapy (MRgRT) for prostate cancer treatments inter-fractional anatomy changes such as bladder and rectum fillings may be corrected by an online adaption of the treatment plan. To clinically implement such complex treatment procedures, however, specific end-to-end tests are required that are able to validate the overall accuracy of all treatment steps from pre-treatment imaging to dose delivery. Approach. In this study, an end-to-end test of a fractionated and online adapted MRgRT prostate irradiation was performed using the so-called ADAM-PETer phantom. The phantom was adapted to perform 3D polymer gel (PG) dosimetry in the prostate and rectum. Furthermore, thermoluminescence detectors (TLDs) were placed at the center and on the surface of the prostate for additional dose measurements as well as for an external dose renormalization of the PG. For the end-to-end test, a total of five online adapted irradiations were applied in sequence with different bladder and rectum fillings, respectively. Main results. A good agreement of measured and planned dose was found represented by high γ -index passing rates ( 3 % / 3 mm criterion) of the PG evaluation of 98.9 % in the prostate and 93.7 % in the rectum. TLDs used for PG renormalization at the center of the prostate showed a deviation of − 2.3 % . Significance. The presented end-to-end test, which allows for 3D dose verification in the prostate and rectum, demonstrates the feasibility and accuracy of fractionated and online-adapted prostate irradiations in presence of inter-fractional anatomy changes. Such tests are of high clinical importance for the commissioning of new image-guided treatment procedures such as online adaptive MRgRT. … (more)
- Is Part Of:
- Physics in medicine & biology. Volume 66:Number 24(2021)
- Journal:
- Physics in medicine & biology
- Issue:
- Volume 66:Number 24(2021)
- Issue Display:
- Volume 66, Issue 24 (2021)
- Year:
- 2021
- Volume:
- 66
- Issue:
- 24
- Issue Sort Value:
- 2021-0066-0024-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-12-16
- Subjects:
- magnetic resonance-guided (adaptive) radiotherapy (MRgRT) -- inter-fractional motion -- fractionated treatment -- end-to-end test -- deformable anthropomorphic pelvis phantom -- 3D gel dosimetry (PAGAT) -- thermoluminescence detectors (TLD)
Biophysics -- Periodicals
Medical physics -- Periodicals
610.153 - Journal URLs:
- http://ioppublishing.org/ ↗
http://iopscience.iop.org/0031-9155 ↗ - DOI:
- 10.1088/1361-6560/ac3e0c ↗
- Languages:
- English
- ISSNs:
- 0031-9155
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 20269.xml