A novel CRT‐IMRT‐combined (Co‐CRIM) planning technique for peripheral lung stereotactic body radiotherapy in pinnacle treatment planning system. Issue 12 (26th October 2021)
- Record Type:
- Journal Article
- Title:
- A novel CRT‐IMRT‐combined (Co‐CRIM) planning technique for peripheral lung stereotactic body radiotherapy in pinnacle treatment planning system. Issue 12 (26th October 2021)
- Main Title:
- A novel CRT‐IMRT‐combined (Co‐CRIM) planning technique for peripheral lung stereotactic body radiotherapy in pinnacle treatment planning system
- Authors:
- Duan, YanHua
Zhou, LiJun
Wang, Hao
Chen, Hua
Gu, HengLe
Shao, Yan
Feng, AiHui
Huang, Ying
Fu, XiaoLong
Yue, Ning Jeff
Ma, Kui
Kong, Qing
Xu, ZhiYong - Abstract:
- Abstract: Objectives: This study attempts to explore a novel peripheral lung stereotactic body radiotherapy (SBRT) planning technique that can balance the pros and cons of three‐dimensional conformal radiotherapy (CRT) and intensity‐modulated radiation therapy (IMRT) / volumetric modulated arc therapy (VMAT). Methods: Treatment plans were retrospectively designed based on CRT, IMRT, VMAT, and the proposed CRT‐IMRT‐combined (Co‐CRIM) techniques using Pinnacle treatment planning system (TPS) for 20 peripheral lung cancer patients. Co‐CRIM used an inverse optimization algorithm available in Pinnacle TPS. To develop a Co‐CRIM plan, the number of segments in each field was limited to one, the minimum segment area was set to the internal target volume (ITV), and the minimum monitor units (MU) of the segment was the quotient of fractional dose divided by twice the number of total fields. The performance of Co‐CRIM was then compared with other techniques. Results: For conformity index (CI), Co‐CRIM performed comparably to IMRT/VMAT but better than CRT. For gradient index (GI), Co‐CRIM was similar to IMRT/VMAT or CRT. For heterogeneity index (HI), Co‐CRIM was comparable to IMRT/VMAT, higher than CRT. The dosimetric results of spinal cord and lung with Co‐CRIM were better than CRT, comparable to IMRT, but inferior to VMAT. The MU resulted from Co‐CRIM was lower than IMRT/VMAT but higher than CRT. For plan verification γ passing rate, Co‐CRIM was higher than IMRT/VMAT, comparable toAbstract: Objectives: This study attempts to explore a novel peripheral lung stereotactic body radiotherapy (SBRT) planning technique that can balance the pros and cons of three‐dimensional conformal radiotherapy (CRT) and intensity‐modulated radiation therapy (IMRT) / volumetric modulated arc therapy (VMAT). Methods: Treatment plans were retrospectively designed based on CRT, IMRT, VMAT, and the proposed CRT‐IMRT‐combined (Co‐CRIM) techniques using Pinnacle treatment planning system (TPS) for 20 peripheral lung cancer patients. Co‐CRIM used an inverse optimization algorithm available in Pinnacle TPS. To develop a Co‐CRIM plan, the number of segments in each field was limited to one, the minimum segment area was set to the internal target volume (ITV), and the minimum monitor units (MU) of the segment was the quotient of fractional dose divided by twice the number of total fields. The performance of Co‐CRIM was then compared with other techniques. Results: For conformity index (CI), Co‐CRIM performed comparably to IMRT/VMAT but better than CRT. For gradient index (GI), Co‐CRIM was similar to IMRT/VMAT or CRT. For heterogeneity index (HI), Co‐CRIM was comparable to IMRT/VMAT, higher than CRT. The dosimetric results of spinal cord and lung with Co‐CRIM were better than CRT, comparable to IMRT, but inferior to VMAT. The MU resulted from Co‐CRIM was lower than IMRT/VMAT but higher than CRT. For plan verification γ passing rate, Co‐CRIM was higher than IMRT/VMAT, comparable to CRT. For planning time, Co‐CRIM was shorter than CRT or VMAT but similar to IMRT. Conclusions: The proposed Co‐CRIM technique on Pinnacle TPS is an effective planning technique for peripheral lung SBRT. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 22:Issue 12(2021)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 22:Issue 12(2021)
- Issue Display:
- Volume 22, Issue 12 (2021)
- Year:
- 2021
- Volume:
- 22
- Issue:
- 12
- Issue Sort Value:
- 2021-0022-0012-0000
- Page Start:
- 97
- Page End:
- 107
- Publication Date:
- 2021-10-26
- Subjects:
- Co‐CRIM -- CRT -- IMRT -- lung SBRT -- planning strategy -- VMAT
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.13461 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 20253.xml