The very large electrode array for retinal stimulation (VLARS)—A concept study. (6th November 2019)
- Record Type:
- Journal Article
- Title:
- The very large electrode array for retinal stimulation (VLARS)—A concept study. (6th November 2019)
- Main Title:
- The very large electrode array for retinal stimulation (VLARS)—A concept study
- Authors:
- Lohmann, Tibor Karl
Haiss, Florent
Schaffrath, Kim
Schnitzler, Anne-Christine
Waschkowski, Florian
Barz, Claudia
van der Meer, Anna-Marina
Werner, Claudia
Johnen, Sandra
Laube, Thomas
Bornfeld, Norbert
Mazinani, Babak Ebrahim
Rößler, Gernot
Mokwa, Wilfried
Walter, Peter - Abstract:
- Abstract: Objective . The restoration of vision in blind patients suffering from degenerative retinal diseases like retinitis pigmentosa may be obtained by local electrical stimulation with retinal implants. In this study, a very large electrode array for retinal stimulation (VLARS) was introduced and tested regarding its safety in implantation and biocompatibility. Further, the array's stimulation capabilities were tested in an acute setting. Approach . The polyimide-based implants have a diameter of 12 mm, cover approximately 110 mm 2 of the retinal surface and carrying 250 iridium oxide coated gold electrodes. The implantation surgery was established in cadaveric porcine eyes. To analyze biocompatibility, ten rabbits were implanted with the VLARS device, and observed for 12 weeks using slit lamp examination, fundus photography, optical coherence tomography (OCT) as well as ultrasound imaging. After enucleation, histological examinations were performed. In acute stimulation experiments, electrodes recorded cortical field potentials upon retinal stimulation in the visual cortex in rabbits. Main results . Implantation studies in rabbits showed that the implantation surgery is safe but difficult. Retinal detachment induced by retinal tears was observed in five animals in varying severity. In five cases, corneal edema reduced the quality of the follow-up examinations. Findings in OCT-imaging and funduscopy suggested that peripheral fixation was insufficient in various animals.Abstract: Objective . The restoration of vision in blind patients suffering from degenerative retinal diseases like retinitis pigmentosa may be obtained by local electrical stimulation with retinal implants. In this study, a very large electrode array for retinal stimulation (VLARS) was introduced and tested regarding its safety in implantation and biocompatibility. Further, the array's stimulation capabilities were tested in an acute setting. Approach . The polyimide-based implants have a diameter of 12 mm, cover approximately 110 mm 2 of the retinal surface and carrying 250 iridium oxide coated gold electrodes. The implantation surgery was established in cadaveric porcine eyes. To analyze biocompatibility, ten rabbits were implanted with the VLARS device, and observed for 12 weeks using slit lamp examination, fundus photography, optical coherence tomography (OCT) as well as ultrasound imaging. After enucleation, histological examinations were performed. In acute stimulation experiments, electrodes recorded cortical field potentials upon retinal stimulation in the visual cortex in rabbits. Main results . Implantation studies in rabbits showed that the implantation surgery is safe but difficult. Retinal detachment induced by retinal tears was observed in five animals in varying severity. In five cases, corneal edema reduced the quality of the follow-up examinations. Findings in OCT-imaging and funduscopy suggested that peripheral fixation was insufficient in various animals. Results of the acute stimulation demonstrated the array's ability to elicit cortical responses. Significance . Overall, it was possible to implant very large epiretinal arrays. On retinal stimulation with the VLARS responses in the visual cortex were recorded. The VLARS device offers the opportunity to restore a much larger field of visual perception when compared to current available retinal implants. … (more)
- Is Part Of:
- Journal of neural engineering. Volume 16:Number 6(2019:Dec.)
- Journal:
- Journal of neural engineering
- Issue:
- Volume 16:Number 6(2019:Dec.)
- Issue Display:
- Volume 16, Issue 6 (2019)
- Year:
- 2019
- Volume:
- 16
- Issue:
- 6
- Issue Sort Value:
- 2019-0016-0006-0000
- Page Start:
- Page End:
- Publication Date:
- 2019-11-06
- Subjects:
- retinal stimulation -- retina implant -- multielectrode array -- cortical activation -- local field potentials -- biocompatibility -- vitreoretinal surgery
Neurosciences -- Periodicals
Biomedical engineering -- Periodicals
612.8 - Journal URLs:
- http://iopscience.iop.org/1741-2552/ ↗
http://ioppublishing.org/ ↗ - DOI:
- 10.1088/1741-2552/ab4113 ↗
- Languages:
- English
- ISSNs:
- 1741-2560
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 20204.xml