Mid-term and long-term follow-up of intracranial aneurysms treated by the p64 Flow Modulation Device: a multicenter experience. (20th July 2016)
- Record Type:
- Journal Article
- Title:
- Mid-term and long-term follow-up of intracranial aneurysms treated by the p64 Flow Modulation Device: a multicenter experience. (20th July 2016)
- Main Title:
- Mid-term and long-term follow-up of intracranial aneurysms treated by the p64 Flow Modulation Device: a multicenter experience
- Authors:
- Briganti, Francesco
Leone, Giuseppe
Ugga, Lorenzo
Marseglia, Mariano
Macera, Antonio
Manto, Andrea
Delehaye, Luigi
Resta, Maurizio
Resta, Mariachiara
Burdi, Nicola
Nuzzi, Nunzio Paolo
Divenuto, Ignazio
Caranci, Ferdinando
Muto, Mario
Solari, Domenico
Cappabianca, Paolo
Maiuri, Francesco - Abstract:
- Abstract : Background: Experience with the endovascular treatment of cerebral aneurysms using the p64 Flow Modulation Device is still limited. This study discusses the results and complications of this new flow diverter device. Methods: 40 patients (30 women, 10 men) with 50 cerebral aneurysms treated in six Italian neurointerventional centers with the p64 Flow Modulation Device between April 2013 and September 2015 were retrospectively reviewed. Results: Complete occlusion was obtained in 44/50 aneurysms (88%) and partial occlusion in 3 (6%). In the other three aneurysms (6%), two cases of asymptomatic in-stent thrombosis and one intraprocedural occlusion of the parent vessel occurred. Technical complications were observed in eight procedures (16%). Permanent morbidity due to acute in-stent thrombosis and consequent ischemic stroke occurred in one patient (2.5%). No delayed aneurysm rupture, subarachnoid or intraparenchymal hemorrhage, or ischemic complications occurred and there were no deaths. Conclusions: Endovascular treatment with the p64 Flow Modulation Device is a safe treatment for unruptured cerebral aneurysms, resulting in a high rate of occlusion. As with other flow diverter devices, we recommend this treatment mainly for large-necked aneurysms of the internal carotid artery siphon. However, endovascular treatment with the p64 device should also be encouraged in difficult cases such as aneurysms of the posterior circulation and beyond the circle of Willis.
- Is Part Of:
- Journal of neurointerventional surgery. Volume 9:Number 1(2017)
- Journal:
- Journal of neurointerventional surgery
- Issue:
- Volume 9:Number 1(2017)
- Issue Display:
- Volume 9, Issue 1 (2017)
- Year:
- 2017
- Volume:
- 9
- Issue:
- 1
- Issue Sort Value:
- 2017-0009-0001-0000
- Page Start:
- 70
- Page End:
- 76
- Publication Date:
- 2016-07-20
- Subjects:
- Aneurysm -- Angiography -- Flow Diverter
Nervous system -- Surgery -- Periodicals
Cerebrovascular disease -- Surgery -- Periodicals
617.48 - Journal URLs:
- http://www.bmj.com/archive ↗
http://jnis.bmj.com/ ↗ - DOI:
- 10.1136/neurintsurg-2016-012502 ↗
- Languages:
- English
- ISSNs:
- 1759-8478
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 20059.xml