A feasibility study of "plan transformation": A novel workaround enabling patient transfer between different Varian machines. (November 2021)
- Record Type:
- Journal Article
- Title:
- A feasibility study of "plan transformation": A novel workaround enabling patient transfer between different Varian machines. (November 2021)
- Main Title:
- A feasibility study of "plan transformation": A novel workaround enabling patient transfer between different Varian machines
- Authors:
- Afsharpour, H.
Nielsen, M.
Jensen, M.
Kaiyum, R.
Cohen, J.
Bozzo, K.
Sankreacha, R. - Abstract:
- Highlights: Plan transformation helps easily transfer plans from Standard to HD MLC Varian units without replanning. Plan transformation is feasible and a better alternative to replanning in terms of time and resource allocation. This process helps us utilize all our resources when a Standard MLC machine goes down for repair. About 25% of our daily workload are transferable from standard to HD unit. Abstract: Purpose: The objective of this work is to explore a methodology to enable plan transfers from Standard MLC (STD) to High Definition MLC (HD) Varian units without replanning. Using "Plan Transformation", we aim to utilize full departmental capacity to disperse patients from STD Varian units to HD units, when the former goes down for repair. Methods: Our centre is equipped with four STD and two HD MLC Varian Truebeam units. Plan transformation is used to transfer patients from STD to HD units when needed. Static or manually segmented fields with Y jaws ≤ 21.6 cm are eligible for plan transformation. The MLC pattern for each field is exported out of Eclipse. An in-house program based on C# is used to convert those patterns from STD to HD. STD and HD MLCs are different in design but dosimetrically comparable for these type of plans. Still to avoid extra uncertainty, transformed plans are recalculated in Eclipse to enable the user to assess the quality of the process. Results: On a given day, more than 25% of appointments on STD units were eligible for transformation. TheHighlights: Plan transformation helps easily transfer plans from Standard to HD MLC Varian units without replanning. Plan transformation is feasible and a better alternative to replanning in terms of time and resource allocation. This process helps us utilize all our resources when a Standard MLC machine goes down for repair. About 25% of our daily workload are transferable from standard to HD unit. Abstract: Purpose: The objective of this work is to explore a methodology to enable plan transfers from Standard MLC (STD) to High Definition MLC (HD) Varian units without replanning. Using "Plan Transformation", we aim to utilize full departmental capacity to disperse patients from STD Varian units to HD units, when the former goes down for repair. Methods: Our centre is equipped with four STD and two HD MLC Varian Truebeam units. Plan transformation is used to transfer patients from STD to HD units when needed. Static or manually segmented fields with Y jaws ≤ 21.6 cm are eligible for plan transformation. The MLC pattern for each field is exported out of Eclipse. An in-house program based on C# is used to convert those patterns from STD to HD. STD and HD MLCs are different in design but dosimetrically comparable for these type of plans. Still to avoid extra uncertainty, transformed plans are recalculated in Eclipse to enable the user to assess the quality of the process. Results: On a given day, more than 25% of appointments on STD units were eligible for transformation. The majority of eligible plans were breast tangent and boost. The delivery and safety of plan transformation was examined directly on a Varian Truebeam linac. A gamma analysis comparison between measured and calculated transformed plans yields a pass-rate of 100% for 3%/3mm, identical to non-transformed plans. Conclusion: Plan transformation is feasible and saves time for planners as it takes less than a quarter the time required for a replan. … (more)
- Is Part Of:
- Physica medica. Volume 91(2021)
- Journal:
- Physica medica
- Issue:
- Volume 91(2021)
- Issue Display:
- Volume 91, Issue 2021 (2021)
- Year:
- 2021
- Volume:
- 91
- Issue:
- 2021
- Issue Sort Value:
- 2021-0091-2021-0000
- Page Start:
- 13
- Page End:
- 17
- Publication Date:
- 2021-11
- Subjects:
- Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2021.10.002 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 20010.xml