Correlation Between Average Segment Width and Gamma Passing Rate as a Function of MLC Position Error in Volumetric Modulated Arc Therapy. (November 2021)
- Record Type:
- Journal Article
- Title:
- Correlation Between Average Segment Width and Gamma Passing Rate as a Function of MLC Position Error in Volumetric Modulated Arc Therapy. (November 2021)
- Main Title:
- Correlation Between Average Segment Width and Gamma Passing Rate as a Function of MLC Position Error in Volumetric Modulated Arc Therapy
- Authors:
- Moon, Young Min
Bae, Sang Il
Han, Moo Jae
Jeon, Wan
Yu, Tosol
Choi, Chul Won
Kim, Jin Young - Abstract:
- Objective: This study analyzed the correlation between the average segment width (ASW) and gamma passing rate according to the multi-leaf collimator (MLC) position error. Method: To evaluate the changes in the gamma passing rate according to the MLC position error, 21 volumetric modulated arc therapy (VMAT) plans were generated using pelvic lymph node metastatic prostate cancer patient's data which is sensitive to MLC position errors as they involve several long, narrow, irregular fields. The ASW for each VMAT plan was calculated using our own code developed using Visual Basic for Applications (VBA). The gamma passing rate of the VMAT plan according to the MLC position error was evaluated using ArcCHECK (Sun Nuclear, Melbourne, FL, USA) while inducing symmetric MLC position errors in 0.25 mm intervals from −1 mm to +1 mm in the infinity medical linear accelerator (Elekta AB, Stockholm, Sweden). Finally, we examined the correlation between the change in the passing rate (γ gradient ) due to the MLC position error and the ASW in VMAT through linear regression analysis using the least squares method. Results: The ASW andγ gradient were found to have a linear correlation according to the MLC position error, and the coefficient of determination was 0.88. For a 1 mm position error of MLC in VMAT, the gamma passing rate improved by approximately 11.9% as the ASW increased by 10 mm. Conclusion: These results are expected to be employed as guidelines to minimize the dose uncertaintyObjective: This study analyzed the correlation between the average segment width (ASW) and gamma passing rate according to the multi-leaf collimator (MLC) position error. Method: To evaluate the changes in the gamma passing rate according to the MLC position error, 21 volumetric modulated arc therapy (VMAT) plans were generated using pelvic lymph node metastatic prostate cancer patient's data which is sensitive to MLC position errors as they involve several long, narrow, irregular fields. The ASW for each VMAT plan was calculated using our own code developed using Visual Basic for Applications (VBA). The gamma passing rate of the VMAT plan according to the MLC position error was evaluated using ArcCHECK (Sun Nuclear, Melbourne, FL, USA) while inducing symmetric MLC position errors in 0.25 mm intervals from −1 mm to +1 mm in the infinity medical linear accelerator (Elekta AB, Stockholm, Sweden). Finally, we examined the correlation between the change in the passing rate (γ gradient ) due to the MLC position error and the ASW in VMAT through linear regression analysis using the least squares method. Results: The ASW andγ gradient were found to have a linear correlation according to the MLC position error, and the coefficient of determination was 0.88. For a 1 mm position error of MLC in VMAT, the gamma passing rate improved by approximately 11.9% as the ASW increased by 10 mm. Conclusion: These results are expected to be employed as guidelines to minimize the dose uncertainty due to MLC position error in VMAT. … (more)
- Is Part Of:
- Technology in cancer research & treatment. Volume 20(2021)
- Journal:
- Technology in cancer research & treatment
- Issue:
- Volume 20(2021)
- Issue Display:
- Volume 20, Issue 2021 (2021)
- Year:
- 2021
- Volume:
- 20
- Issue:
- 2021
- Issue Sort Value:
- 2021-0020-2021-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-11
- Subjects:
- VMAT -- average segment width -- MLC position error -- gamma passing rate -- radiation therapy
Oncology -- Periodicals
Cancer -- Diagnosis -- Periodicals
Cancer -- Treatment -- Technological innovations -- Periodicals
616.994 - Journal URLs:
- http://tct.sagepub.com/ ↗
http://www.tcrt.org ↗
http://www.sagepub.com ↗ - DOI:
- 10.1177/15330338211059937 ↗
- Languages:
- English
- ISSNs:
- 1533-0346
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 19940.xml