Dosimetric Evaluation of the QFix kVueTM Calypso Couch Top. (28th April 2021)
- Record Type:
- Journal Article
- Title:
- Dosimetric Evaluation of the QFix kVueTM Calypso Couch Top. (28th April 2021)
- Main Title:
- Dosimetric Evaluation of the QFix kVueTM Calypso Couch Top
- Authors:
- Hou, Lingtong
Zhang, Huiqin
Sun, Xiaomei
Liu, Qianqian
Chen, Tingfeng
Liu, Yong
Jiang, Xiaodong
Yao, Shengyu - Abstract:
- Purpose: To evaluate the dosimetric accuracy of the default couch model of the QFix kVue TM Calypso couch top in the treatment planning system. Methods: With the gantry 180°, field size 20 × 20 cm, 6 MV, we measured the depth dose, off-axis dose, and dose plane of different depths in the phantom with the couch rails in and out, respectively. Isocenter doses at different angles were also obtained. The results were compared to the doses calculated using the default couch top model and the real scanned couch top model. Then we revised the default model according to the measured results. Results: With "Rails In, " the depth dose, off-axis dose, and dose plane of the default couch top model had a big difference with the dose of the real scanned couch top model and the measured result. The dose of the real scanned couch top model was much closer to the measured result, but in the region of the rail edge, the difference was still significant. With "Rails Out, " there was a minor difference between the measured result, the dose of the default couch top model and the real scanned couch top model. The difference between the measurement and the default couch top model became very small after being revised. Conclusions: It is better to avoid the beam angle passing through the couch rails in treatment plans, or you should revise the parameter of the QFix kVue TM Calypso couch top model based on the measured results, and verify the treatment plan before clinical practice.
- Is Part Of:
- Technology in cancer research & treatment. Volume 20(2021)
- Journal:
- Technology in cancer research & treatment
- Issue:
- Volume 20(2021)
- Issue Display:
- Volume 20, Issue 2021 (2021)
- Year:
- 2021
- Volume:
- 20
- Issue:
- 2021
- Issue Sort Value:
- 2021-0020-2021-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-04-28
- Subjects:
- Calypso -- couch rails -- default model -- dosimetric accuracy -- model revision
Oncology -- Periodicals
Cancer -- Diagnosis -- Periodicals
Cancer -- Treatment -- Technological innovations -- Periodicals
616.994 - Journal URLs:
- http://tct.sagepub.com/ ↗
http://www.tcrt.org ↗
http://www.sagepub.com ↗ - DOI:
- 10.1177/15330338211011964 ↗
- Languages:
- English
- ISSNs:
- 1533-0346
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File: