Force distribution within the frame of self-expanding transcatheter aortic valve: Insights from in-vivo finite element analysis. (9th November 2021)
- Record Type:
- Journal Article
- Title:
- Force distribution within the frame of self-expanding transcatheter aortic valve: Insights from in-vivo finite element analysis. (9th November 2021)
- Main Title:
- Force distribution within the frame of self-expanding transcatheter aortic valve: Insights from in-vivo finite element analysis
- Authors:
- Xiong, Tian-Yuan
Stoppani, Elisa
De Beule, Matthieu
Chen, Fei
Li, Yi-Jian
Liao, Yan-Biao
Feng, Yuan
de Jaegere, Peter
Chen, Mao - Abstract:
- Abstract: We sought to assess the amount and distribution of force on the valve frame after transcatheter aortic valve replacement (TAVR) via patient-specific computer simulation. Patients successfully treated with the self-expanding Venus A-Valve and multislice computed tomography (MSCT) pre- and post-TAVR were retrospectively included. Patient-specific finite element models of the aortic root and prosthesis were constructed. The force (in Newton) on the valve frame was derived at every 3 mm from the inflow and at every 22.5° on each level. Twenty patients of whom 10 had bicuspid aortic valve (BAV) were analyzed. The total force on the frame was 74.9 N in median (interquartile range 24.0). The maximal force was observed at level 5 that corresponds with the nadir of the bioprosthetic leaflets and was 9.9 (7.1) N in all patients, 10.3 (6.6) N in BAV and 9.7 (9.2) N for patients with tricuspid aortic valve (TAV). The level of maximal force located higher from the native annulus in BAV and TAV patients (8.8 [4.8] vs. 1.8 [7.4] mm). The area of the valve frame at the level of maximal force decreased from 437.4 (239.7) mm 2 at the annulus to 377.6 (114.3) mm 2 in BAV, but increased from 397.5 (114.3) mm 2 at the annulus to 406.7 (108.9) mm 2 in TAV. The maximum force on the bioprosthetic valve frame is located at the plane of the nadir of the bioprosthetic leaflets. It remains to be elucidated whether this may be associated with bioprosthetic frame and leaflet integrity and/orAbstract: We sought to assess the amount and distribution of force on the valve frame after transcatheter aortic valve replacement (TAVR) via patient-specific computer simulation. Patients successfully treated with the self-expanding Venus A-Valve and multislice computed tomography (MSCT) pre- and post-TAVR were retrospectively included. Patient-specific finite element models of the aortic root and prosthesis were constructed. The force (in Newton) on the valve frame was derived at every 3 mm from the inflow and at every 22.5° on each level. Twenty patients of whom 10 had bicuspid aortic valve (BAV) were analyzed. The total force on the frame was 74.9 N in median (interquartile range 24.0). The maximal force was observed at level 5 that corresponds with the nadir of the bioprosthetic leaflets and was 9.9 (7.1) N in all patients, 10.3 (6.6) N in BAV and 9.7 (9.2) N for patients with tricuspid aortic valve (TAV). The level of maximal force located higher from the native annulus in BAV and TAV patients (8.8 [4.8] vs. 1.8 [7.4] mm). The area of the valve frame at the level of maximal force decreased from 437.4 (239.7) mm 2 at the annulus to 377.6 (114.3) mm 2 in BAV, but increased from 397.5 (114.3) mm 2 at the annulus to 406.7 (108.9) mm 2 in TAV. The maximum force on the bioprosthetic valve frame is located at the plane of the nadir of the bioprosthetic leaflets. It remains to be elucidated whether this may be associated with bioprosthetic frame and leaflet integrity and/or function. … (more)
- Is Part Of:
- Journal of biomechanics. Volume 128(2021)
- Journal:
- Journal of biomechanics
- Issue:
- Volume 128(2021)
- Issue Display:
- Volume 128, Issue 2021 (2021)
- Year:
- 2021
- Volume:
- 128
- Issue:
- 2021
- Issue Sort Value:
- 2021-0128-2021-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-11-09
- Subjects:
- Transcatheter aortic valve replacement -- Finite element analysis -- Computer simulation -- Force
Animal mechanics -- Periodicals
Biomechanics -- Periodicals
Biomechanics -- Periodicals
Mécanique animale -- Périodiques
Biomécanique -- Périodiques
Electronic journals
571.4305 - Journal URLs:
- http://www.sciencedirect.com/science/journal/00219290 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/00219290 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/00219290 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jbiomech.2021.110804 ↗
- Languages:
- English
- ISSNs:
- 0021-9290
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4953.600000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 19909.xml