A novel partial resuscitative endovascular balloon aortic occlusion device that can be deployed in zone 1 for more than 2 hours with minimal provider titration. Issue 3 (22nd March 2021)
- Record Type:
- Journal Article
- Title:
- A novel partial resuscitative endovascular balloon aortic occlusion device that can be deployed in zone 1 for more than 2 hours with minimal provider titration. Issue 3 (22nd March 2021)
- Main Title:
- A novel partial resuscitative endovascular balloon aortic occlusion device that can be deployed in zone 1 for more than 2 hours with minimal provider titration
- Authors:
- Kemp, Michael T.
Wakam, Glenn K.
Williams, Aaron M.
Biesterveld, Ben E.
O'Connell, Rachel L.
Vercruysse, Claire A.
Chtraklin, Kiril
Russo, Rachel M.
Alam, Hasan B. - Abstract:
- Abstract : BACKGROUND: Hemorrhage is a leading cause of mortality in trauma. Resuscitative endovascular balloon occlusion of the aorta (REBOA) can control hemorrhage, but distal ischemia, subsequent reperfusion injury, and the need for frequent balloon titration remain problems. Improved device design can allow for partial REBOA (pREBOA) that may provide hemorrhage control while also perfusing distally without need for significant provider titration. METHODS: Female Yorkshire swine (N = 10) were subjected to 40% hemorrhagic shock for 1 hour (mean arterial pressure [MAP], 28–32 mm Hg). Animals were then randomized to either complete aortic occlusion (ER-REBOA) or partial occlusion (novel pREBOA-PRO) without frequent provider titration or distal MAP targets. Detection of a trace distal waveform determined partial occlusion in the pREBOA-PRO arm. After 2 hours of zone 1 occlusion, the hemorrhaged whole blood was returned. After 50% autotransfusion, the balloon was deflated over a 10-minute period. Following transfusion, the animals were survived for 2 hours while receiving resuscitation based on objective targets: lactated Ringer's fluid boluses (goal central venous pressure, ≥ 6 mm Hg), a norepinephrine infusion (goal MAP, 55–60 mm Hg), and acid-base correction (goal pH, >7.2). Hemodynamic variables, arterial lactate, lactate dehydrogenase, aspartate aminotransferase, and creatinine levels were measured. RESULTS: All animals survived throughout the experiment, with similarAbstract : BACKGROUND: Hemorrhage is a leading cause of mortality in trauma. Resuscitative endovascular balloon occlusion of the aorta (REBOA) can control hemorrhage, but distal ischemia, subsequent reperfusion injury, and the need for frequent balloon titration remain problems. Improved device design can allow for partial REBOA (pREBOA) that may provide hemorrhage control while also perfusing distally without need for significant provider titration. METHODS: Female Yorkshire swine (N = 10) were subjected to 40% hemorrhagic shock for 1 hour (mean arterial pressure [MAP], 28–32 mm Hg). Animals were then randomized to either complete aortic occlusion (ER-REBOA) or partial occlusion (novel pREBOA-PRO) without frequent provider titration or distal MAP targets. Detection of a trace distal waveform determined partial occlusion in the pREBOA-PRO arm. After 2 hours of zone 1 occlusion, the hemorrhaged whole blood was returned. After 50% autotransfusion, the balloon was deflated over a 10-minute period. Following transfusion, the animals were survived for 2 hours while receiving resuscitation based on objective targets: lactated Ringer's fluid boluses (goal central venous pressure, ≥ 6 mm Hg), a norepinephrine infusion (goal MAP, 55–60 mm Hg), and acid-base correction (goal pH, >7.2). Hemodynamic variables, arterial lactate, lactate dehydrogenase, aspartate aminotransferase, and creatinine levels were measured. RESULTS: All animals survived throughout the experiment, with similar increase in proximal MAPs in both groups. Animals that underwent partial occlusion had slightly higher distal MAPs. At the end of the experiment, the partial occlusion group had lower end levels of serum lactate ( p = 0.006), lactate dehydrogenase ( p = 0.0004) and aspartate aminotransferase ( p = 0.004). Animals that underwent partial occlusion required less norepinephrine ( p = 0.002), less bicarbonate administration ( p = 0.006), and less fluid resuscitation ( p = 0.042). CONCLUSION: Improved design for pREBOA can decrease the degree of distal ischemia and reperfusion injury compared with complete aortic occlusion, while providing a similar increase in proximal MAPs. This can allow pREBOA zone-1 deployment for longer periods without the need for significant balloon titration. … (more)
- Is Part Of:
- Journal of trauma and acute care surgery. Volume 90:Issue 3(2021)
- Journal:
- Journal of trauma and acute care surgery
- Issue:
- Volume 90:Issue 3(2021)
- Issue Display:
- Volume 90, Issue 3 (2021)
- Year:
- 2021
- Volume:
- 90
- Issue:
- 3
- Issue Sort Value:
- 2021-0090-0003-0000
- Page Start:
- 426
- Page End:
- 433
- Publication Date:
- 2021-03-22
- Subjects:
- REBOA -- partial aortic occlusion -- titration -- resuscitation -- hemorrhagic shock -- organ injury
Surgical intensive care -- Periodicals
Surgical emergencies -- Periodicals
Wounds and injuries -- Surgery -- Periodicals
617.026 - Journal URLs:
- http://journals.lww.com/jtrauma/pages/default.aspx ↗
http://ovidsp.tx.ovid.com/sp-3.5.0b/ovidweb.cgi?&S=NEIKFPIGHGDDBOHLNCALMDIBGLDKAA00&Browse=Toc+Children%7cNO%7cS.sh.2697_1327404888_15.2697_1327404888_27.2697_1327404888_28%7c273%7c50 ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/TA.0000000000003042 ↗
- Languages:
- English
- ISSNs:
- 2163-0755
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5070.510500
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 19813.xml