Distinguishing atrial fibrillation from sinus rhythm using commercial pulse detection systems: The non-interventional BAYathlon study. (May 2021)
- Record Type:
- Journal Article
- Title:
- Distinguishing atrial fibrillation from sinus rhythm using commercial pulse detection systems: The non-interventional BAYathlon study. (May 2021)
- Main Title:
- Distinguishing atrial fibrillation from sinus rhythm using commercial pulse detection systems: The non-interventional BAYathlon study
- Authors:
- Müller, Christian
Hengstmann, Ulf
Fuchs, Michael
Kirchner, Martin
Kleinjung, Frank
Mathis, Harald
Martin, Stephan
Bläse, Ingo
Perings, Stefan - Abstract:
- Objective: Early diagnosis of atrial fibrillation (AFib) is a priority for stroke prevention. We sought to test four commercial pulse detection systems (CPDSs) for ability to distinguish AFib from normal sinus rhythm using a published algorithm (Zhou et al., PLoS One 2015;10:e0136544), compared with visual diagnosis by electrocardiogram inspection. Methods: BAYathlon was a prospective, non-interventional, single-centre study. Adult cardiology patients with documented AFib or sinus rhythm who were due to have a routine 5-min electrocardiogram were randomized to undergo a parallel 5-min pulse assessment with a Polar V800, eMotion Faros 360, TomTom heart rate monitor, or Adidas miCoach Smart Run. Results: 144 patients (73 with AFib, 71 with sinus rhythm (based on electrocardiograms); median age: 73 years; 53.5% male) were analysed. Algorithm sensitivities (primary endpoint) and specificities for AFib when applied to CPDS recordings were 93.3% and 94.1% with the Polar V800, 90.0% and 84.2% with the eMotion Faros 360, and 0% and 100% with the other CPDSs (analysis period: 127 heart rate signals + 2 min). When applied to routine electrocardiograms, the algorithm correctly detected AFib in 71/73 patients. Different analysis periods (127 heart rate signals +1 or 3 min) only slightly changed the sensitivities with the Polar V800 and eMotion Faros 360 and had no effect on the sensitivities with the other CPDSs. Conclusion: AFib screening using the applied algorithm is feasible withObjective: Early diagnosis of atrial fibrillation (AFib) is a priority for stroke prevention. We sought to test four commercial pulse detection systems (CPDSs) for ability to distinguish AFib from normal sinus rhythm using a published algorithm (Zhou et al., PLoS One 2015;10:e0136544), compared with visual diagnosis by electrocardiogram inspection. Methods: BAYathlon was a prospective, non-interventional, single-centre study. Adult cardiology patients with documented AFib or sinus rhythm who were due to have a routine 5-min electrocardiogram were randomized to undergo a parallel 5-min pulse assessment with a Polar V800, eMotion Faros 360, TomTom heart rate monitor, or Adidas miCoach Smart Run. Results: 144 patients (73 with AFib, 71 with sinus rhythm (based on electrocardiograms); median age: 73 years; 53.5% male) were analysed. Algorithm sensitivities (primary endpoint) and specificities for AFib when applied to CPDS recordings were 93.3% and 94.1% with the Polar V800, 90.0% and 84.2% with the eMotion Faros 360, and 0% and 100% with the other CPDSs (analysis period: 127 heart rate signals + 2 min). When applied to routine electrocardiograms, the algorithm correctly detected AFib in 71/73 patients. Different analysis periods (127 heart rate signals +1 or 3 min) only slightly changed the sensitivities with the Polar V800 and eMotion Faros 360 and had no effect on the sensitivities with the other CPDSs. Conclusion: AFib screening using the applied algorithm is feasible with the Polar V800 and eMotion Faros 360 (which provide RR interval data) but not with the other CPDSs (which provide pre-processed heart rate time series). ClinicalTrials.gov identifier: NCT02875106 … (more)
- Is Part Of:
- Digital health. Volume 7(2021)
- Journal:
- Digital health
- Issue:
- Volume 7(2021)
- Issue Display:
- Volume 7, Issue 2021 (2021)
- Year:
- 2021
- Volume:
- 7
- Issue:
- 2021
- Issue Sort Value:
- 2021-0007-2021-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-05
- Subjects:
- Atrial fibrillation -- electrocardiography -- algorithms -- sensitivity -- observational study -- sinus rhythm
Medical care -- Data processing -- Periodicals
Medical informatics -- Periodicals
362.10285 - Journal URLs:
- http://www.uk.sagepub.com/home.nav ↗
http://dhj.sagepub.com/ ↗ - DOI:
- 10.1177/20552076211019620 ↗
- Languages:
- English
- ISSNs:
- 2055-2076
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 19787.xml