024 Validation of tissue characterisation and vulnerable plaque classification using virtual histology IVUS (VH-IVUS) against human post-mortem histology. (16th May 2012)
- Record Type:
- Journal Article
- Title:
- 024 Validation of tissue characterisation and vulnerable plaque classification using virtual histology IVUS (VH-IVUS) against human post-mortem histology. (16th May 2012)
- Main Title:
- 024 Validation of tissue characterisation and vulnerable plaque classification using virtual histology IVUS (VH-IVUS) against human post-mortem histology
- Authors:
- Obaid, D R
Calvert, P A
Goddard, M
Bennett, M R - Abstract:
- Abstract : Introduction: VH-IVUS is increasingly used in clinical trials to classify vulnerable thin-capped fibroatheroma (TCFA). However, VH-IVUS has not been validated for classifying coronary plaque against the gold standard of human post-mortem histology. Methods: Ten human coronary arteries were obtained at post-mortem. VH-IVUS examination was performed in a pressure-perfused system. The arteries were then fixed, stained and sectioned giving 72 co-registered 400 μm segments. Slides from each segment were examined by a histopathologist to determine the presence of calcification, necrotic core and to classify any plaque present. A fibroatheroma was considered a TCFA if fibrous cap thickness was <65 μm. VH-IVUS analysis was performed on each segment to compare tissue characterisation and plaque classification by an operator blinded to the histology. To explore a possible link between VH-IVUS detection of calcification and necrotic core artefact we also performed VH-IVUS of coronary segments in 20 patients pre- and immediately post-stenting to determine if there was a relationship between increasing calcium (simulated by stent struts) and necrotic core. Results: VH-IVUS was excellent at determining the presence of atherosclerotic plaque (sensitivity-100%, specificity-93%) and calcification (sensitivity-96%, specificity-90%) (Abstract 024 figure 1 ). Necrotic core was detected with a sensitivity of 100% but a specificity of 40%. For classification of a fibroatheroma byAbstract : Introduction: VH-IVUS is increasingly used in clinical trials to classify vulnerable thin-capped fibroatheroma (TCFA). However, VH-IVUS has not been validated for classifying coronary plaque against the gold standard of human post-mortem histology. Methods: Ten human coronary arteries were obtained at post-mortem. VH-IVUS examination was performed in a pressure-perfused system. The arteries were then fixed, stained and sectioned giving 72 co-registered 400 μm segments. Slides from each segment were examined by a histopathologist to determine the presence of calcification, necrotic core and to classify any plaque present. A fibroatheroma was considered a TCFA if fibrous cap thickness was <65 μm. VH-IVUS analysis was performed on each segment to compare tissue characterisation and plaque classification by an operator blinded to the histology. To explore a possible link between VH-IVUS detection of calcification and necrotic core artefact we also performed VH-IVUS of coronary segments in 20 patients pre- and immediately post-stenting to determine if there was a relationship between increasing calcium (simulated by stent struts) and necrotic core. Results: VH-IVUS was excellent at determining the presence of atherosclerotic plaque (sensitivity-100%, specificity-93%) and calcification (sensitivity-96%, specificity-90%) (Abstract 024 figure 1 ). Necrotic core was detected with a sensitivity of 100% but a specificity of 40%. For classification of a fibroatheroma by VH-IVUS (confluent NC >10% of plaque area for three consecutive frames) sensitivity was 87% and specificity-74%. Of the 24 segments wrongly attributed to contain necrotic core by VH-IVUS, 92% also contained calcium (67% contained >10% calcium). Stent struts are mistaken by VH-IVUS as calcification, surrounded by necrotic core not present on histology (Abstract 024 figure 2 ). Analysis of arterial segments pre- and post-stenting showed a linear correlation between increasing calcium (stent struts) with increases in necrotic core (r 2 =0.61) suggesting some necrotic core surrounding calcium maybe artefact. VH-IVUS distinguished TCFA (fibroatheroma with core in contact with lumen) with sensitivity-71% and specificity-74%. As 65 μm is beyond the spatial resolution of VH-IVUS we repeated the analysis with caps <200 μm considered thin and the sensitivity of VH-IVUS to detect TCFA rose to 80%, specificity-76% (Abstract 024 table 1 ). Conclusion: In comparison with histology, VH-IVUS reliably discriminated calcified plaque and detected necrotic core (classifying fibroatheromas) with good sensitivity. However, the specificity to detect necrotic core dropped in the presence of dense calcification due to necrotic core artefact. The sensitivity and specificity to detect TCFA was also limited by VH-IVUS spatial resolution. … (more)
- Is Part Of:
- Heart. Volume 98(2012)Supplement 1
- Journal:
- Heart
- Issue:
- Volume 98(2012)Supplement 1
- Issue Display:
- Volume 98, Issue 1 (2012)
- Year:
- 2012
- Volume:
- 98
- Issue:
- 1
- Issue Sort Value:
- 2012-0098-0001-0000
- Page Start:
- A16
- Page End:
- A16
- Publication Date:
- 2012-05-16
- Subjects:
- VH-IVUS -- histology -- TCFA
Heart -- Diseases -- Treatment -- Periodicals
Cardiology -- Periodicals
616.12 - Journal URLs:
- http://www.bmj.com/archive ↗
http://heart.bmj.com ↗
http://www.heartjnl.com ↗ - DOI:
- 10.1136/heartjnl-2012-301877b.24 ↗
- Languages:
- English
- ISSNs:
- 1355-6037
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 19731.xml