51 Outcomes with single-coil versus dual-coil implantable cardioverter-defibrillators: a meta-analysis. (5th June 2017)
- Record Type:
- Journal Article
- Title:
- 51 Outcomes with single-coil versus dual-coil implantable cardioverter-defibrillators: a meta-analysis. (5th June 2017)
- Main Title:
- 51 Outcomes with single-coil versus dual-coil implantable cardioverter-defibrillators: a meta-analysis
- Authors:
- Sunderland, Nicholas
Kaura, Amit
Murgatroyd, Francis
Dhillon, Para
Scott, Paul - Abstract:
- Abstract : Aims: Dual-coil implantable cardioverter defibrillator (ICD) leads have traditionally been used over single-coil leads due to concerns regarding high defibrillation thresholds (DFT) and consequent poor shock efficacy. However, accumulating evidence suggests that this position may be unfounded and that dual-coil leads may also be associated with higher complication rates during lead extraction. This meta-analysis collates data comparing dual- and single-coil ICD leads. Methods and results: Electronic databases were systematically searched for randomised controlled trials (RCT) and non-randomised studies comparing single-coil and dual-coil leads. The mean differences in DFT and summary estimates of the odds-ratio (OR) for first-shock efficacy and the hazard-ratio (HR) for all-cause mortality were calculated using random effects models. Eighteen studies including a total of 1 38 124 patients were identified. Dual-coil leads were associated with a lower DFT compared to single coil leads (mean difference –0.83J; 95% confidence interval [CI] –1.39–−0.27; p=0.004). There was no difference in the first-shock success rate with dual-coil compared to single-coil leads (OR 0.74; 95% CI 0.45–1.21; p=0.22). There was a significantly lower risk of all-cause mortality associated with single-coil leads (HR 0.91; 95% CI 0.86–0.95; p<0.0001). Conclusion: This meta-analysis suggests that single-coil leads have a marginally higher DFT but that this may be clinically insignificant asAbstract : Aims: Dual-coil implantable cardioverter defibrillator (ICD) leads have traditionally been used over single-coil leads due to concerns regarding high defibrillation thresholds (DFT) and consequent poor shock efficacy. However, accumulating evidence suggests that this position may be unfounded and that dual-coil leads may also be associated with higher complication rates during lead extraction. This meta-analysis collates data comparing dual- and single-coil ICD leads. Methods and results: Electronic databases were systematically searched for randomised controlled trials (RCT) and non-randomised studies comparing single-coil and dual-coil leads. The mean differences in DFT and summary estimates of the odds-ratio (OR) for first-shock efficacy and the hazard-ratio (HR) for all-cause mortality were calculated using random effects models. Eighteen studies including a total of 1 38 124 patients were identified. Dual-coil leads were associated with a lower DFT compared to single coil leads (mean difference –0.83J; 95% confidence interval [CI] –1.39–−0.27; p=0.004). There was no difference in the first-shock success rate with dual-coil compared to single-coil leads (OR 0.74; 95% CI 0.45–1.21; p=0.22). There was a significantly lower risk of all-cause mortality associated with single-coil leads (HR 0.91; 95% CI 0.86–0.95; p<0.0001). Conclusion: This meta-analysis suggests that single-coil leads have a marginally higher DFT but that this may be clinically insignificant as there appears to be no difference in first-shock efficacy when compared to dual-coil leads. The mortality benefit with single-coil leads most likely represents patient selection bias. Given the increased risk and complexity of extracting dual-coil leads, centres should strongly consider single-coil ICD leads as the lead of choice for routine new left-sided ICD implants. … (more)
- Is Part Of:
- Heart. Volume 103(2017)Supplement 5
- Journal:
- Heart
- Issue:
- Volume 103(2017)Supplement 5
- Issue Display:
- Volume 103, Issue 5 (2017)
- Year:
- 2017
- Volume:
- 103
- Issue:
- 5
- Issue Sort Value:
- 2017-0103-0005-0000
- Page Start:
- A40
- Page End:
- A40
- Publication Date:
- 2017-06-05
- Subjects:
- Implantable Cardioverter-Defibrillators -- Meta-analysis -- Single-coil leads
Heart -- Diseases -- Treatment -- Periodicals
Cardiology -- Periodicals
616.12 - Journal URLs:
- http://www.bmj.com/archive ↗
http://heart.bmj.com ↗
http://www.heartjnl.com ↗ - DOI:
- 10.1136/heartjnl-2017-311726.50 ↗
- Languages:
- English
- ISSNs:
- 1355-6037
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 19675.xml