A retrospective comparison of organ dose and effective dose in percutaneous vertebroplasty performed under CT guidance or using a fixed C-arm with a flat-panel detector. (August 2021)
- Record Type:
- Journal Article
- Title:
- A retrospective comparison of organ dose and effective dose in percutaneous vertebroplasty performed under CT guidance or using a fixed C-arm with a flat-panel detector. (August 2021)
- Main Title:
- A retrospective comparison of organ dose and effective dose in percutaneous vertebroplasty performed under CT guidance or using a fixed C-arm with a flat-panel detector
- Authors:
- Greffier, Joël
Hamard, Aymeric
Berny, Laure
Snene, Fehmi
Perolat, Romain
Larbi, Ahmed
Viala, Pierre
Dabli, Djamel
Beregi, Jean-Paul - Abstract:
- Highlights: Effective dose was significantly higher for PVP performed under CT guidance. Organ doses were significantly higher with CT-scan than with fixed C-arm, except for oesophagus. From a dosimetry point of view, PVP should preferentially be performed with the fixed C-arm. Abstract: Purpose: To compare the organ-dose and effective-dose (E) delivered to the patient during percutaneous vertebroplasty (PVP) of one thoracic or lumbar vertebra performed under CT guidance or using a fixed C-arm. Methods: Consecutive adult patients undergoing PVP of one vertebra under CT-guidance, with optimized protocol and training of physicians, or using a fixed C-arm were retrospectively included from January 2016 to June 2017. Organ-doses were computed on 16 organs using CT Expo 2.4 software for the CT procedures and PCXMC 2.0 for the fixed C-arm procedures. E was also computed with both software. Dosimetric values per anatomic locations for all procedures were compared using the paired Mann-Whitney-Wilcoxon test. Results: In total, 73 patients were analysed (27 men and 46 women, mean age 78 ± 10 years) among whom 35 (48%) underwent PVP under CT guidance and 38 (52%) PVP using a fixed C-arm. The median E was 11.31 [6.54; 15.82] mSv for all PVPs performed under CT guidance and 5.58 [3.33; 8.71] mSv for fixed C-arm and the differences was significant (p<0.001). For lumbar PVP, the organ doses of stomach, liver and colon were significantly higher with CT-scan than with the fixed C-arm: 97%Highlights: Effective dose was significantly higher for PVP performed under CT guidance. Organ doses were significantly higher with CT-scan than with fixed C-arm, except for oesophagus. From a dosimetry point of view, PVP should preferentially be performed with the fixed C-arm. Abstract: Purpose: To compare the organ-dose and effective-dose (E) delivered to the patient during percutaneous vertebroplasty (PVP) of one thoracic or lumbar vertebra performed under CT guidance or using a fixed C-arm. Methods: Consecutive adult patients undergoing PVP of one vertebra under CT-guidance, with optimized protocol and training of physicians, or using a fixed C-arm were retrospectively included from January 2016 to June 2017. Organ-doses were computed on 16 organs using CT Expo 2.4 software for the CT procedures and PCXMC 2.0 for the fixed C-arm procedures. E was also computed with both software. Dosimetric values per anatomic locations for all procedures were compared using the paired Mann-Whitney-Wilcoxon test. Results: In total, 73 patients were analysed (27 men and 46 women, mean age 78 ± 10 years) among whom 35 (48%) underwent PVP under CT guidance and 38 (52%) PVP using a fixed C-arm. The median E was 11.31 [6.54; 15.82] mSv for all PVPs performed under CT guidance and 5.58 [3.33; 8.71] mSv for fixed C-arm and the differences was significant (p<0.001). For lumbar PVP, the organ doses of stomach, liver and colon were significantly higher with CT-scan than with the fixed C-arm: 97% (p=0.02); 21% (p=0.099) and 375% (p=0.002), respectively. For thoracic PVP, the lung organ dose was significantly higher with CT-scan than with the fixed C-arm (127%; p<0.001) and the oesophagus organ doses were not significantly different (p = 0.626). Conclusion: This study showed that the E and the organ dose on directly exposed organs were both higher for PVP performed under CT-guidance than with the fixed C-arm. … (more)
- Is Part Of:
- Physica medica. Volume 88(2021)
- Journal:
- Physica medica
- Issue:
- Volume 88(2021)
- Issue Display:
- Volume 88, Issue 2021 (2021)
- Year:
- 2021
- Volume:
- 88
- Issue:
- 2021
- Issue Sort Value:
- 2021-0088-2021-0000
- Page Start:
- 235
- Page End:
- 241
- Publication Date:
- 2021-08
- Subjects:
- AK Air Kerma -- CTDIvol Volume CT dose index -- DAP Dose area product -- DLP Dose length product -- E Effective dose -- PVP Percutaneous vertebroplasty
Vertebroplasty -- Multidetector computed tomography -- Interventional radiology -- Organ dose -- Effective dose
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2021.07.002 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 19603.xml