Accuracy of the volume and pressure displays of high frequency oscillators. Issue 2 (1st March 2004)
- Record Type:
- Journal Article
- Title:
- Accuracy of the volume and pressure displays of high frequency oscillators. Issue 2 (1st March 2004)
- Main Title:
- Accuracy of the volume and pressure displays of high frequency oscillators
- Authors:
- Leipälä, J A
Iwasaki, S
Milner, A
Greenough, A - Abstract:
- Abstract : Objective: To determine the effect of frequency on the accuracy of volume and pressure displays of high frequency oscillators. Methods: The effect of frequency on the displayed volume of the Stephanie, Dräger Babylog 8000 Plus, and SLE 5000 oscillators was assessed. A sine wave pump delivered a constant tidal volume at frequencies of 5–15 Hz to the patient manifold of the oscillators. The displayed volumes at each frequency were compared with the delivered volume. The effect of frequency on displayed pressure was assessed by connecting the oscillator's patient manifold to a lung model; three types of oscillator were studied (SensorMedics 3100A, SLE 5000, and Stephanie). Airway pressure was measured from the manifold using a pressure transducer and non-compliant tubing; the pressure measuring system had a flat frequency response to 30 Hz. Results: The SLE 5000 volume display overread the delivered volume (by about 5%), but was not affected by frequency. At 5 Hz, the Dräger Babylog 8000 Plus and the Stephanie underread the delivered volume (by about 20%). Increasing frequency resulted in a greater discrepancy between the delivered and displayed volume with the Stephanie, but a smaller discrepancy with the Dräger Babylog 8000 Plus. Altering frequency had a small effect (maximum difference 6%) on the relation between the displayed and delivered pressure for all three oscillators. Conclusion: Frequency affects the accuracy of displayed volumes and, to a lesser extent,Abstract : Objective: To determine the effect of frequency on the accuracy of volume and pressure displays of high frequency oscillators. Methods: The effect of frequency on the displayed volume of the Stephanie, Dräger Babylog 8000 Plus, and SLE 5000 oscillators was assessed. A sine wave pump delivered a constant tidal volume at frequencies of 5–15 Hz to the patient manifold of the oscillators. The displayed volumes at each frequency were compared with the delivered volume. The effect of frequency on displayed pressure was assessed by connecting the oscillator's patient manifold to a lung model; three types of oscillator were studied (SensorMedics 3100A, SLE 5000, and Stephanie). Airway pressure was measured from the manifold using a pressure transducer and non-compliant tubing; the pressure measuring system had a flat frequency response to 30 Hz. Results: The SLE 5000 volume display overread the delivered volume (by about 5%), but was not affected by frequency. At 5 Hz, the Dräger Babylog 8000 Plus and the Stephanie underread the delivered volume (by about 20%). Increasing frequency resulted in a greater discrepancy between the delivered and displayed volume with the Stephanie, but a smaller discrepancy with the Dräger Babylog 8000 Plus. Altering frequency had a small effect (maximum difference 6%) on the relation between the displayed and delivered pressure for all three oscillators. Conclusion: Frequency affects the accuracy of displayed volumes and, to a lesser extent, displayed pressures of high frequency oscillators. The results emphasise that data displayed by new devices should not be uncritically accepted. … (more)
- Is Part Of:
- Archives of disease in childhood. Volume 89:Issue 2(2004)
- Journal:
- Archives of disease in childhood
- Issue:
- Volume 89:Issue 2(2004)
- Issue Display:
- Volume 89, Issue 2 (2004)
- Year:
- 2004
- Volume:
- 89
- Issue:
- 2
- Issue Sort Value:
- 2004-0089-0002-0000
- Page Start:
- F174
- Page End:
- F176
- Publication Date:
- 2004-03-01
- Subjects:
- high frequency oscillation -- prematurity -- tidal volume
Infants -- Diseases -- Periodicals
Newborn infants -- Diseases -- Periodicals
Fetus -- Diseases -- Periodicals
618.920105 - Journal URLs:
- http://fn.bmjjournals.com ↗
http://www.bmj.com/archive ↗ - DOI:
- 10.1136/adc.2002.023937 ↗
- Languages:
- English
- ISSNs:
- 1359-2998
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 19538.xml