Identification of elemental weight fraction and mass density of humanoid tissue-equivalent materials using dual energy computed tomography. (July 2017)
- Record Type:
- Journal Article
- Title:
- Identification of elemental weight fraction and mass density of humanoid tissue-equivalent materials using dual energy computed tomography. (July 2017)
- Main Title:
- Identification of elemental weight fraction and mass density of humanoid tissue-equivalent materials using dual energy computed tomography
- Authors:
- Shimomura, Kohei
Araki, Fujio
Kono, Yuki
Asai, Yoshiyuki
Murakami, Takamichi
Hyodo, Tomoko
Okumura, Masahiko
Matsumoto, Kenji
Monzen, Hajime
Nishimura, Yasumasa - Abstract:
- Highlights: A novel method to identity for physical parameters of humanoid phantoms is proposed. Elemental weight fraction and mass density are calculated by least-squares technique. It is possible to identify physical parameters more accurate than previous data. Abstract: The purpose of this study was to obtain the fraction by weight of the elemental composition and mass density of a humanoid tissue phantom that includes lung tissue, soft tissue, and bone tissue, by using dual energy computed tomography (DECT). The fraction by weight and the mass density for tissue-equivalent materials were calculated by means of a least-squares method with a linear attenuation coefficient, using monochromatic photon energies of 10–140 keV, as obtained from DECT. The accuracy of calculated values for the fractions by weight of H (hydrogen), C (carbon), N (nitrogen), and O (oxygen) as verified by comparing the values with those that were analyzed using the combustion technique. The fraction by weight for other elements was confirmed by comparing with the analyzed values by means of energy dispersive photon spectroscopy. The calculated mass densities for each tissue were compared with those that were obtained by dividing the weight by volume. The calculated values of the fraction by weight that were obtained by means of DECT had differences of 1.9%, 9.2%, 6.6%, 7.8%, 0.8%, and 0.2% at a maximum for H, C, N, O, P (phosphorus), and Ca (calcium), respectively, from the reference values analyzedHighlights: A novel method to identity for physical parameters of humanoid phantoms is proposed. Elemental weight fraction and mass density are calculated by least-squares technique. It is possible to identify physical parameters more accurate than previous data. Abstract: The purpose of this study was to obtain the fraction by weight of the elemental composition and mass density of a humanoid tissue phantom that includes lung tissue, soft tissue, and bone tissue, by using dual energy computed tomography (DECT). The fraction by weight and the mass density for tissue-equivalent materials were calculated by means of a least-squares method with a linear attenuation coefficient, using monochromatic photon energies of 10–140 keV, as obtained from DECT. The accuracy of calculated values for the fractions by weight of H (hydrogen), C (carbon), N (nitrogen), and O (oxygen) as verified by comparing the values with those that were analyzed using the combustion technique. The fraction by weight for other elements was confirmed by comparing with the analyzed values by means of energy dispersive photon spectroscopy. The calculated mass densities for each tissue were compared with those that were obtained by dividing the weight by volume. The calculated values of the fraction by weight that were obtained by means of DECT had differences of 1.9%, 9.2%, 6.6%, 7.8%, 0.8%, and 0.2% at a maximum for H, C, N, O, P (phosphorus), and Ca (calcium), respectively, from the reference values analyzed by the combustion technique and energy dispersive photon spectroscopy. The difference in the mass density for tissue was 0.011 g/cm 3 at a maximum. This study demonstrated the fraction by weight and the mass density of the humanoid tissue-equivalent materials that were calculated by means of DECT were expected high accuracy. … (more)
- Is Part Of:
- Physica medica. Volume 39(2017)
- Journal:
- Physica medica
- Issue:
- Volume 39(2017)
- Issue Display:
- Volume 39, Issue 2017 (2017)
- Year:
- 2017
- Volume:
- 39
- Issue:
- 2017
- Issue Sort Value:
- 2017-0039-2017-0000
- Page Start:
- 59
- Page End:
- 66
- Publication Date:
- 2017-07
- Subjects:
- Dual energy CT -- Elemental weight fraction -- Mass density -- Monte Carlo simulation
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2017.05.060 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 19121.xml