Early results of the Axium MicroFX for Endovascular Repair of IntraCranial Aneurysm (AMERICA) study: a multicenter prospective observational registry. (11th September 2013)
- Record Type:
- Journal Article
- Title:
- Early results of the Axium MicroFX for Endovascular Repair of IntraCranial Aneurysm (AMERICA) study: a multicenter prospective observational registry. (11th September 2013)
- Main Title:
- Early results of the Axium MicroFX for Endovascular Repair of IntraCranial Aneurysm (AMERICA) study: a multicenter prospective observational registry
- Authors:
- Fargen, Kyle M
Blackburn, Spiros
Carpenter, Jeffrey S
Jabbour, Pascal
Mack, William J
Rai, Ansaar T
Siddiqui, Adnan H
Turner, Raymond D
Mocco, J - Abstract:
- Abstract : Background: Axium MicroFX coils contain polymer (polyglycolic/polylactic acid, PGLA) microfilaments designed to significantly impact intra-aneurysmal flow and to encourage aneurysm thrombosis. To provide preliminary evaluation of the safety and suggested efficacy of the MicroFX design, we performed a multicenter prospective single-arm trial, Axium M icroFX for E ndovascular R epair of I ntraC ranial Aneurysm study (AMERICA). Methods: AMERICA is a prospective multicenter 100 aneurysm observational study evaluating the safety and efficacy of Axium MicroFX PGLA coils. Enrollment was started in April 2010 and completed in October 2012. Results: 99 patients underwent treatment for 100 aneurysms at 13 centers. Mean age was 60.2 years, most were women (72%) and 18% of patients had previously undergone treatment for a separate aneurysm. 22% of patients underwent treatment after acute aneurysmal subarachnoid hemorrhage (SAH). Of these patients, all were Hunt and Hess grade 1–3. Pre-procedure modified Rankin score (mRS) was 0–2 in 92% of patients. The majority of aneurysms were anterior circulation (86%), with the most common aneurysm locations being the anterior communicating artery (23%) followed by the supraclinoid internal carotid artery and posterior communicating artery (18% each). The mean maximum diameter was 6.5 mm. Axium MicroFX coils could be placed in all but one treatment (99%). Raymond grade at conclusion of coiling was I (complete) in 52%, II (dog ear orAbstract : Background: Axium MicroFX coils contain polymer (polyglycolic/polylactic acid, PGLA) microfilaments designed to significantly impact intra-aneurysmal flow and to encourage aneurysm thrombosis. To provide preliminary evaluation of the safety and suggested efficacy of the MicroFX design, we performed a multicenter prospective single-arm trial, Axium M icroFX for E ndovascular R epair of I ntraC ranial Aneurysm study (AMERICA). Methods: AMERICA is a prospective multicenter 100 aneurysm observational study evaluating the safety and efficacy of Axium MicroFX PGLA coils. Enrollment was started in April 2010 and completed in October 2012. Results: 99 patients underwent treatment for 100 aneurysms at 13 centers. Mean age was 60.2 years, most were women (72%) and 18% of patients had previously undergone treatment for a separate aneurysm. 22% of patients underwent treatment after acute aneurysmal subarachnoid hemorrhage (SAH). Of these patients, all were Hunt and Hess grade 1–3. Pre-procedure modified Rankin score (mRS) was 0–2 in 92% of patients. The majority of aneurysms were anterior circulation (86%), with the most common aneurysm locations being the anterior communicating artery (23%) followed by the supraclinoid internal carotid artery and posterior communicating artery (18% each). The mean maximum diameter was 6.5 mm. Axium MicroFX coils could be placed in all but one treatment (99%). Raymond grade at conclusion of coiling was I (complete) in 52%, II (dog ear or residual neck) in 33% and III (residual aneurysm) in 15%. Discharge mRS was significantly worse in patients with SAH (62% mRS 0–2) compared with electively treated aneurysms (mRS 0–2 in 94%, p<001). Major events were uncommon (6% thromboembolic events, 3% intraoperative vessel or aneurysm rupture) and device-related adverse events (AE) were rare (2% of cases). Conclusions: This prospective study of Axium MicroFX coils demonstrates excellent aneurysm occlusion rates. 52% of aneurysms were completely occluded post-procedure. Within the ruptured aneurysm group, post-procedure occlusion rates were 63.6%. Major AE rates were consistent with historical data. … (more)
- Is Part Of:
- Journal of neurointerventional surgery. Volume 6:Number 7(2014)
- Journal:
- Journal of neurointerventional surgery
- Issue:
- Volume 6:Number 7(2014)
- Issue Display:
- Volume 6, Issue 7 (2014)
- Year:
- 2014
- Volume:
- 6
- Issue:
- 7
- Issue Sort Value:
- 2014-0006-0007-0000
- Page Start:
- 495
- Page End:
- 499
- Publication Date:
- 2013-09-11
- Subjects:
- Aneurysm -- Coil -- Subarachnoid
Nervous system -- Surgery -- Periodicals
Cerebrovascular disease -- Surgery -- Periodicals
617.48 - Journal URLs:
- http://www.bmj.com/archive ↗
http://jnis.bmj.com/ ↗ - DOI:
- 10.1136/neurintsurg-2013-010887 ↗
- Languages:
- English
- ISSNs:
- 1759-8478
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 19067.xml