Validation of the Microlife BP B3 AFIB upper arm blood pressure monitor in adults and adolescents according to the ANSI/AAMI/ISO 81060-2: 2019 protocol. Issue 4 (August 2021)
- Record Type:
- Journal Article
- Title:
- Validation of the Microlife BP B3 AFIB upper arm blood pressure monitor in adults and adolescents according to the ANSI/AAMI/ISO 81060-2: 2019 protocol. Issue 4 (August 2021)
- Main Title:
- Validation of the Microlife BP B3 AFIB upper arm blood pressure monitor in adults and adolescents according to the ANSI/AAMI/ISO 81060-2
- Authors:
- Beime, Beate
Bramlage, Carsten
Krüger, Ralf
Deutsch, Cornelia
van Mark, Gesine
Bramlage, Peter
Botta, Beate - Abstract:
- Abstract : Objective: Aim of this study was to validate the Microlife BP B3 AFIB/enterprise resource planning (ERP) No: BP3KT1-3 N blood pressure (BP) monitor according to the American National Standards Institute (ANSI)/Association for the Advancement of Medical Instrumentation (AAMI)/International Organization for Standardization (ISO) 81060–2:2019 in adolescents and adults from a general population. Methods: BP measurements on the upper arm were performed in 85 subjects (age range 12–88 years), using the Microlife BP B3 AFIB and a standard mercury reference sphygmomanometer. Results: A total of 255 valid BP comparisons were performed for the present validation analysis. The mean ± SD difference between the test and the reference device was 0.70 ± 7.05 mmHg for SBP (pass criterion ⩽5 mmHg) and −0.85 ± 4.70 mmHg for DBP (pass criterion ⩽5 mmHg) with the SD below the required value of ⩽8 mmHg. The mean ± SD of the intraindividual differences between the test and the reference device was 0.70 ± 5.87 mmHg for SBP (pass criterion for the SD ⩽6.90 mmHg) and −0.85 ± 4.19 mmHg for DBP (pass criterion for the SD ⩽6.88 mmHg). Conclusion: The Microlife BP B3 AFIB/ERP No: BP3KT1-3 N has passed the criteria of the ANSI/AAMI/ISO 81060–2:2019 protocol and can be recommended for home BP measurements in adolescents and adults.
- Is Part Of:
- Blood pressure monitoring. Volume 26:Issue 4(2021)
- Journal:
- Blood pressure monitoring
- Issue:
- Volume 26:Issue 4(2021)
- Issue Display:
- Volume 26, Issue 4 (2021)
- Year:
- 2021
- Volume:
- 26
- Issue:
- 4
- Issue Sort Value:
- 2021-0026-0004-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-08
- Subjects:
- American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization -- blood pressure monitor -- Microlife BP B3 AFIB -- validation
Blood pressure -- Measurement -- Periodicals
Blood Pressure Determination -- Periodicals
Blood Pressure Monitoring, Ambulatory -- Periodicals
Blood Pressure Monitors -- Periodicals
Hypertension -- Periodicals
612.140287 - Journal URLs:
- http://gateway.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&PAGE=toc&D=ovft&AN=00126097-000000000-00000&NEWS=N ↗
http://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=toc&D=yrovft&AN=00126097-000000000-00000 ↗
http://www.bpmonitoring.com/ ↗
http://journals.lww.com/pages/default.aspx ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1097/MBP.0000000000000530 ↗
- Languages:
- English
- ISSNs:
- 1359-5237
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 2113.035500
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 18948.xml