The impact of coil shape design on angiographic occlusion, packing density and coil mass uniformity in aneurysm embolization: an in vitro study. (24th January 2011)
- Record Type:
- Journal Article
- Title:
- The impact of coil shape design on angiographic occlusion, packing density and coil mass uniformity in aneurysm embolization: an in vitro study. (24th January 2011)
- Main Title:
- The impact of coil shape design on angiographic occlusion, packing density and coil mass uniformity in aneurysm embolization: an in vitro study
- Authors:
- Mehra, Manik
Hurley, Michael C
Gounis, Matthew J
King, Robert M
Shaibani, Ali
Dabus, Guilherme
Labdag, Fatimaezzahra E
Levy, Elad I
Bendok, Bernard R - Abstract:
- Abstract : Objective: To investigate the impact of coil design on the distribution of the coil mass in a controlled in vitro experiment. A secondary objective was to study the relationship between angiographic occlusion, packing density and coil mass uniformity. Methods: Seven silicon side wall aneurysm models were embolized in each arm with a different coil design under fluoroscopic guidance. Packing density calculations and scoring of the angiographic occlusion were made. The models were embedded in epoxy and sectioned through the aneurysm neck and the dome. The sections were imaged and processed to derive the fractional surface area of coil material over the coil free area, a quantitative endpoint representing the percentage of the given region of interest consumed by coils. The SD of the surface area fractions is inversely proportional to the uniformity of coil distribution. Results: The novel triangular primary wind design of the Deltapaq achieved a more homogenous distribution of coils within the aneurysm dome compared with the helical and complex microcoil systems (p=0.018). The packing density achieved by the Deltapaq (39.1±1.6%) was significantly higher than the complex (35.2±2.8%) and the helical (32.2±3.3%) coils. Angiographically, aneurysms coiled by the Deltapaq were more likely to obtain a Raymond score of class I. Conclusion: Evaluation of emerging coil technologies with respect to treatment durability may be well served by an assessment of their uniformity ofAbstract : Objective: To investigate the impact of coil design on the distribution of the coil mass in a controlled in vitro experiment. A secondary objective was to study the relationship between angiographic occlusion, packing density and coil mass uniformity. Methods: Seven silicon side wall aneurysm models were embolized in each arm with a different coil design under fluoroscopic guidance. Packing density calculations and scoring of the angiographic occlusion were made. The models were embedded in epoxy and sectioned through the aneurysm neck and the dome. The sections were imaged and processed to derive the fractional surface area of coil material over the coil free area, a quantitative endpoint representing the percentage of the given region of interest consumed by coils. The SD of the surface area fractions is inversely proportional to the uniformity of coil distribution. Results: The novel triangular primary wind design of the Deltapaq achieved a more homogenous distribution of coils within the aneurysm dome compared with the helical and complex microcoil systems (p=0.018). The packing density achieved by the Deltapaq (39.1±1.6%) was significantly higher than the complex (35.2±2.8%) and the helical (32.2±3.3%) coils. Angiographically, aneurysms coiled by the Deltapaq were more likely to obtain a Raymond score of class I. Conclusion: Evaluation of emerging coil technologies with respect to treatment durability may be well served by an assessment of their uniformity of distribution within an aneurysm in addition of the traditional packing density and angiographic occlusion scoring methods. … (more)
- Is Part Of:
- Journal of neurointerventional surgery. Volume 3:Number 2(2011)
- Journal:
- Journal of neurointerventional surgery
- Issue:
- Volume 3:Number 2(2011)
- Issue Display:
- Volume 3, Issue 2 (2011)
- Year:
- 2011
- Volume:
- 3
- Issue:
- 2
- Issue Sort Value:
- 2011-0003-0002-0000
- Page Start:
- 131
- Page End:
- 136
- Publication Date:
- 2011-01-24
- Subjects:
- Nervous system -- Surgery -- Periodicals
Cerebrovascular disease -- Surgery -- Periodicals
617.48 - Journal URLs:
- http://www.bmj.com/archive ↗
http://jnis.bmj.com/ ↗ - DOI:
- 10.1136/jnis.2010.004390 ↗
- Languages:
- English
- ISSNs:
- 1759-8478
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 18898.xml