Monitoring sterilization of dressings in high-vacuum pressure-steam sterilizers. Issue 6 (November 1961)
- Record Type:
- Journal Article
- Title:
- Monitoring sterilization of dressings in high-vacuum pressure-steam sterilizers. Issue 6 (November 1961)
- Main Title:
- Monitoring sterilization of dressings in high-vacuum pressure-steam sterilizers
- Authors:
- Fallon, R. J.
- Abstract:
- Abstract : Undetected air leaks in a high-vacuum sterilizer can defeat the sterilization process. For this reason the maintenance and supervision of these sterilizers needs expert engineering care. The importance of controlling the high-vacuum sterilization process from the centre of a load rather than the chamber drain is emphasized, and a method of detecting any failure in a high-vacuum process is described. The dynamics of the high-vacuum sterilization process are discussed and the possible disadvantages of an integrator are noted. The wisdom of using a temperature of 134°C. for three minutes is questioned.
- Is Part Of:
- Journal of clinical pathology. Volume 14:Issue 6(1961)
- Journal:
- Journal of clinical pathology
- Issue:
- Volume 14:Issue 6(1961)
- Issue Display:
- Volume 14, Issue 6 (1961)
- Year:
- 1961
- Volume:
- 14
- Issue:
- 6
- Issue Sort Value:
- 1961-0014-0006-0000
- Page Start:
- 666
- Page End:
- 669
- Publication Date:
- 1961-11
- Subjects:
- Pathology -- Periodicals
Pathology, Molecular -- Periodicals
616.0705 - Journal URLs:
- http://jcp.bmjjournals.com ↗
http://jcp.bmjjournals.com/content/by/year ↗
http://www.pubmedcentral.nih.gov/tocrender.fcgi?journal=162&action=archive ↗
http://www.bmj.com/archive ↗ - DOI:
- 10.1136/jcp.14.6.666 ↗
- Languages:
- English
- ISSNs:
- 0021-9746
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 18842.xml