Air Embolism during Haemodialysis. Issue 5766 (10th July 1971)
- Record Type:
- Journal Article
- Title:
- Air Embolism during Haemodialysis. Issue 5766 (10th July 1971)
- Main Title:
- Air Embolism during Haemodialysis
- Authors:
- Ward, M. K.
Shadforth, M.
Hill, A. V. L.
Kerr, D. N. S. - Abstract:
- Abstract : Two fatal and five non-fatal episodes of air embolism occurred in the course of 14, 000 haemodialyses. Air entered the circuit through the infusion bottle, the heparin syringe or line, or the blood pump insert. Emergency treatment with posture, oxygen, dextran infusion, and dexamethasone was apparently successful in reversing some of the manifestations. Prevention depends partly on better design of equipment and partly on the incorporation in the haemodialysis circuit of devices which detect air and prevent it from reaching the patient. Such monitors are not in general use in the United Kingdom and those currently available are far from infallible. We describe an alternative system incorporating a floating valve in the bubble trap.
- Is Part Of:
- BMJ. Volume 3:Issue 5766(1971)
- Journal:
- BMJ
- Issue:
- Volume 3:Issue 5766(1971)
- Issue Display:
- Volume 3, Issue 5766 (1971)
- Year:
- 1971
- Volume:
- 3
- Issue:
- 5766
- Issue Sort Value:
- 1971-0003-5766-0000
- Page Start:
- 74
- Page End:
- 78
- Publication Date:
- 1971-07-10
- Subjects:
- Medicine -- Periodicals
Medicine -- Periodicals
Medicine
Periodicals
610 - Journal URLs:
- http://www.bmj.com/archive ↗
http://www.jstor.org/journals/09598138.html ↗
http://www.ncbi.nlm.nih.gov/pmc/journals/3/ ↗
http://www.bmj.com/bmj/ ↗
http://www.bmj.com/archive ↗ - DOI:
- 10.1136/bmj.3.5766.74 ↗
- Languages:
- English
- ISSNs:
- 0007-1447
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 18750.xml