MR safety watchdog for active catheters: Wireless impedance control with real‐time feedback. Issue 2 (21st January 2020)
- Record Type:
- Journal Article
- Title:
- MR safety watchdog for active catheters: Wireless impedance control with real‐time feedback. Issue 2 (21st January 2020)
- Main Title:
- MR safety watchdog for active catheters: Wireless impedance control with real‐time feedback
- Authors:
- Özen, Ali Caglar
Silemek, Berk
Lottner, Thomas
Atalar, Ergin
Bock, Michael - Abstract:
- Abstract : Purpose: To dynamically minimize radiofrequency (RF)‐induced heating of an active catheter through an automatic change of the termination impedance. Methods: A prototype wireless module was designed that modifies the input impedance of an active catheter to keep the temperature rise during MRI below a threshold, ΔTmax . The wireless module (MR safety watchdog; MRsWD) measures the local temperature at the catheter tip using either a built‐in thermistor or external data from a fiber‐optical thermometer. It automatically changes the catheter input impedance until the temperature rise during MRI is minimized. If ΔTmax is exceeded, RF transmission is blocked by a feedback system. Results: The thermistor and fiber‐optical thermometer provided consistent temperature data in a phantom experiment. During MRI, the MRsWD was able to reduce the maximum temperature rise by 25% when operated in real‐time feedback mode. Conclusion: This study demonstrates the technical feasibility of an MRsWD as an alternative or complementary approach to reduce RF‐induced heating of active interventional devices. The automatic MRsWD can reduce heating using direct temperature measurements at the tip of the catheter. Given that temperature measurements are intrinsically slow, for a clinical implementation, a faster feedback parameter would be required such as the RF currents along the catheter or scattered electric fields at the tip.
- Is Part Of:
- Magnetic resonance in medicine. Volume 84:Issue 2(2020)
- Journal:
- Magnetic resonance in medicine
- Issue:
- Volume 84:Issue 2(2020)
- Issue Display:
- Volume 84, Issue 2 (2020)
- Year:
- 2020
- Volume:
- 84
- Issue:
- 2
- Issue Sort Value:
- 2020-0084-0002-0000
- Page Start:
- 1048
- Page End:
- 1060
- Publication Date:
- 2020-01-21
- Subjects:
- active catheter -- active implantable medical devices -- bluetooth low energy -- interventional MRI -- MR safety -- radio frequency induced heating
Nuclear magnetic resonance -- Periodicals
Electron paramagnetic resonance -- Periodicals
616.07548 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2594 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/mrm.28153 ↗
- Languages:
- English
- ISSNs:
- 0740-3194
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5337.798000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 18710.xml