Automatic 3D Monte-Carlo-based secondary dose calculation for online verification of 1.5 T magnetic resonance imaging guided radiotherapy. (July 2021)
- Record Type:
- Journal Article
- Title:
- Automatic 3D Monte-Carlo-based secondary dose calculation for online verification of 1.5 T magnetic resonance imaging guided radiotherapy. (July 2021)
- Main Title:
- Automatic 3D Monte-Carlo-based secondary dose calculation for online verification of 1.5 T magnetic resonance imaging guided radiotherapy
- Authors:
- Nachbar, Marcel
Mönnich, David
Dohm, Oliver
Friedlein, Melissa
Zips, Daniel
Thorwarth, Daniela - Abstract:
- Highlights: First implementation of an independent 3D-secondary dose calculation (3D-SDC). Validation of the 3D-SDC solution using patient plans and experimental plan QA. Online SDC of central targets is feasible with a median calculation time of 1:23 min. Peripheral targets with small beam numbers need alternative validation strategies. Abstract: Background and purpose: Hybrid magnetic resonance linear accelerator (MR-Linac) systems represent a novel technology for online adaptive radiotherapy. 3D secondary dose calculation (SDC) of online adapted plans is required to assure patient safety. Currently, no 3D-SDC solution is available for 1.5T MR-Linac systems. Therefore, the aim of this project was to develop and validate a method for online automatic 3D-SDC for adaptive MR-Linac treatments. Materials and methods: An accelerator head model was designed for an 1.5T MR-Linac system, neglecting the magnetic field. The use of this model for online 3D-SDC of MR-Linac plans was validated in a three-step process: (1) comparison to measured beam data, (2) investigation of performance and limitations in a planning phantom and (3) clinical validation using n = 100 patient plans from different tumor entities, comparing the developed 3D-SDC with experimental plan QA. Results: The developed model showed median gamma passing rates compared to MR-Linac base data of 84.7%, 100% and 99.1% for crossplane, inplane and depth-dose-profiles, respectively. Comparison of 3D-SDC and full doseHighlights: First implementation of an independent 3D-secondary dose calculation (3D-SDC). Validation of the 3D-SDC solution using patient plans and experimental plan QA. Online SDC of central targets is feasible with a median calculation time of 1:23 min. Peripheral targets with small beam numbers need alternative validation strategies. Abstract: Background and purpose: Hybrid magnetic resonance linear accelerator (MR-Linac) systems represent a novel technology for online adaptive radiotherapy. 3D secondary dose calculation (SDC) of online adapted plans is required to assure patient safety. Currently, no 3D-SDC solution is available for 1.5T MR-Linac systems. Therefore, the aim of this project was to develop and validate a method for online automatic 3D-SDC for adaptive MR-Linac treatments. Materials and methods: An accelerator head model was designed for an 1.5T MR-Linac system, neglecting the magnetic field. The use of this model for online 3D-SDC of MR-Linac plans was validated in a three-step process: (1) comparison to measured beam data, (2) investigation of performance and limitations in a planning phantom and (3) clinical validation using n = 100 patient plans from different tumor entities, comparing the developed 3D-SDC with experimental plan QA. Results: The developed model showed median gamma passing rates compared to MR-Linac base data of 84.7%, 100% and 99.1% for crossplane, inplane and depth-dose-profiles, respectively. Comparison of 3D-SDC and full dose calculation in a planning phantom revealed that with ⩾ 5 beams gamma passing rates > 95% can be achieved for central target locations. With a median calculation time of 1:23 min, 3D-SDC of online adapted clinical MR-Linac plans demonstrated a median gamma passing rate of 98.9% compared to full dose calculation, whereas experimental plan QA reached 99.5%. Conclusion: Here, we describe the first technical 3D-SDC solution for online adaptive MR-guided radiotherapy. For clinical situations with peripheral targets and a small number of beams additional verification appears necessary. Further improvement may include 3D-SDC with consideration of the magnetic field. … (more)
- Is Part Of:
- Physics and imaging in radiation oncology. Volume 19(2021)
- Journal:
- Physics and imaging in radiation oncology
- Issue:
- Volume 19(2021)
- Issue Display:
- Volume 19, Issue 2021 (2021)
- Year:
- 2021
- Volume:
- 19
- Issue:
- 2021
- Issue Sort Value:
- 2021-0019-2021-0000
- Page Start:
- 6
- Page End:
- 12
- Publication Date:
- 2021-07
- Subjects:
- MR-guided radiotherapy -- MR-Linac -- Online secondary dose calculation -- Adaptive radiotherapy -- Online plan quality assurance
Radiotherapy -- Periodicals
Radiation dosimetry -- Periodicals
Cancer -- Imaging -- Periodicals
Oncology -- Periodicals
615.842 - Journal URLs:
- http://www.sciencedirect.com/ ↗
https://www.journals.elsevier.com/physics-and-imaging-in-radiation-oncology/ ↗ - DOI:
- 10.1016/j.phro.2021.05.002 ↗
- Languages:
- English
- ISSNs:
- 2405-6316
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File: