Repeatability and intermediate precision of a mass concentration calibration system. Issue 6 (3rd June 2019)
- Record Type:
- Journal Article
- Title:
- Repeatability and intermediate precision of a mass concentration calibration system. Issue 6 (3rd June 2019)
- Main Title:
- Repeatability and intermediate precision of a mass concentration calibration system
- Authors:
- Titosky, Jordan
Momenimovahed, Ali
Corbin, Joel
Thomson, Kevin
Smallwood, Greg
Olfert, Jason S. - Abstract:
- Abstract: Many aerosol instruments require calibration to make accurate measurements. A centrifugal particle mass analyzer (CPMA) and aerosol electrometer can be used to calibrate aerosol instruments that measure mass concentration. To understand the sources of uncertainty in the calibration method, two CPMA-electrometer systems were tested to measure the repeatability and intermediate precision of the system, where the repeatability is the standard deviation of several measurements using the same system over a short period of time, and the intermediate precision is the standard deviation of several measurements using different instruments with different calibrations over a long period of time. It was found that the repeatability of the CPMA and the aerosol electrometer were both 0.8%, while the intermediate precision was 1.3% and 2.2%, respectively. The intermediate precision of the aerosol electrometers determined here compares well with a broader study by national metrology institutes which determined an intermediate precision of ∼1.7%. By propagation of uncertainty, it is expected that a CPMA-electrometer system would have repeatability of 1.1% and an intermediate precision of ∼2.1%. This compares favorably to thermal-optical analysis methods which aim to measure black carbon mass concentrations for instrument calibration, which have a repeatability in the range of 8.5–20% and reproducibility in the range of 20–26% for elemental carbon. Thus, the CPMA-electrometer methodAbstract: Many aerosol instruments require calibration to make accurate measurements. A centrifugal particle mass analyzer (CPMA) and aerosol electrometer can be used to calibrate aerosol instruments that measure mass concentration. To understand the sources of uncertainty in the calibration method, two CPMA-electrometer systems were tested to measure the repeatability and intermediate precision of the system, where the repeatability is the standard deviation of several measurements using the same system over a short period of time, and the intermediate precision is the standard deviation of several measurements using different instruments with different calibrations over a long period of time. It was found that the repeatability of the CPMA and the aerosol electrometer were both 0.8%, while the intermediate precision was 1.3% and 2.2%, respectively. The intermediate precision of the aerosol electrometers determined here compares well with a broader study by national metrology institutes which determined an intermediate precision of ∼1.7%. By propagation of uncertainty, it is expected that a CPMA-electrometer system would have repeatability of 1.1% and an intermediate precision of ∼2.1%. This compares favorably to thermal-optical analysis methods which aim to measure black carbon mass concentrations for instrument calibration, which have a repeatability in the range of 8.5–20% and reproducibility in the range of 20–26% for elemental carbon. Thus, the CPMA-electrometer method may be a good alternative to existing instrument calibration procedures. Copyright © 2019 American Association for Aerosol Research … (more)
- Is Part Of:
- Aerosol science and technology. Volume 53:Issue 6(2019)
- Journal:
- Aerosol science and technology
- Issue:
- Volume 53:Issue 6(2019)
- Issue Display:
- Volume 53, Issue 6 (2019)
- Year:
- 2019
- Volume:
- 53
- Issue:
- 6
- Issue Sort Value:
- 2019-0053-0006-0000
- Page Start:
- 701
- Page End:
- 711
- Publication Date:
- 2019-06-03
- Subjects:
- Jingkun Jiang
Aerosols -- Periodicals
Aerosol Propellants -- Periodicals
Aerosols -- Periodicals
660.294515 - Journal URLs:
- http://www.tandfonline.com/loi/uast20#.VkNQFJUnyig ↗
http://www.tandfonline.com/ ↗ - DOI:
- 10.1080/02786826.2019.1592103 ↗
- Languages:
- English
- ISSNs:
- 0278-6826
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0729.835400
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 18591.xml