PTU-048 The Effect of the Colonoscope Magnetic Imaging Device (Scopeguide ®) on Implantable Cardiac Devices. (4th June 2013)
- Record Type:
- Journal Article
- Title:
- PTU-048 The Effect of the Colonoscope Magnetic Imaging Device (Scopeguide ®) on Implantable Cardiac Devices. (4th June 2013)
- Main Title:
- PTU-048 The Effect of the Colonoscope Magnetic Imaging Device (Scopeguide ®) on Implantable Cardiac Devices
- Authors:
- Corbett, G D
Lim, Y C
Pugh, P
Cameron, E - Abstract:
- Abstract : Introduction: Use of the colonoscope magnetic imaging device (ScopeGuide®, Olympus) is currently contraindicated by the manufacturer for patients with implantable cardiac devices. This group of patients is increasing every year, as is the number of colonoscopies being performed. This is the first study which examines the safety of ScopeGuide® in patients with permanent pacemakers (PPMs), implantable cardioverter defibrillators (ICDs) and cardiac resynchronisation therapy (CRT). Methods: Before any patient experiments, ex-vivo proof of concept studies were performed. Non-implanted devices were exposed to the electromagnetic field (EMF) of ScopeGuide® and monitoring demonstrated no evidence of interference or change in settings. Following this, ethical approval was obtained and consecutive patients attending device cheque clinics were prospectively recruited and exposed to ScopeGuide® EMF. After an initial device cheque, patients were attached to a device programmer and underwent continuous external cardiac monitoring. A colonoscope was placed curled on the abdomen over their clothing with patients lying supine to simulate the EMF generated during colonoscopy. The colonoscope was connected to ScopeGuide®. After 2 minutes of observation to assess for any interference, ScopeGuide® was disconnected and devices were checked for any change in settings. Results: 143 patients were invited to participate and of these 70 were recruited to take part in the study (46 male, 24Abstract : Introduction: Use of the colonoscope magnetic imaging device (ScopeGuide®, Olympus) is currently contraindicated by the manufacturer for patients with implantable cardiac devices. This group of patients is increasing every year, as is the number of colonoscopies being performed. This is the first study which examines the safety of ScopeGuide® in patients with permanent pacemakers (PPMs), implantable cardioverter defibrillators (ICDs) and cardiac resynchronisation therapy (CRT). Methods: Before any patient experiments, ex-vivo proof of concept studies were performed. Non-implanted devices were exposed to the electromagnetic field (EMF) of ScopeGuide® and monitoring demonstrated no evidence of interference or change in settings. Following this, ethical approval was obtained and consecutive patients attending device cheque clinics were prospectively recruited and exposed to ScopeGuide® EMF. After an initial device cheque, patients were attached to a device programmer and underwent continuous external cardiac monitoring. A colonoscope was placed curled on the abdomen over their clothing with patients lying supine to simulate the EMF generated during colonoscopy. The colonoscope was connected to ScopeGuide®. After 2 minutes of observation to assess for any interference, ScopeGuide® was disconnected and devices were checked for any change in settings. Results: 143 patients were invited to participate and of these 70 were recruited to take part in the study (46 male, 24 females, mean age 68.5 years SD 10.3). Device type, indication and manufacturer are summarised in Table 1. No evidence of interference was seen on device leads nor was any change in programming detected following exposure to ScopeGuide® EMF in any patient. Conclusion: The ScopeGuide® does not cause any interference or change in settings and appears to be safe to use in patients with implantable cardiac devices. Disclosure of Interest: None Declared … (more)
- Is Part Of:
- Gut. Volume 62(2013)Supplement 1
- Journal:
- Gut
- Issue:
- Volume 62(2013)Supplement 1
- Issue Display:
- Volume 62, Issue 1 (2013)
- Year:
- 2013
- Volume:
- 62
- Issue:
- 1
- Issue Sort Value:
- 2013-0062-0001-0000
- Page Start:
- A63
- Page End:
- A63
- Publication Date:
- 2013-06-04
- Subjects:
- Gastroenterology -- Periodicals
616.33 - Journal URLs:
- http://gut.bmjjournals.com ↗
http://www.bmj.com/archive ↗ - DOI:
- 10.1136/gutjnl-2013-304907.140 ↗
- Languages:
- English
- ISSNs:
- 0017-5749
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 18580.xml