Analytical evaluation of direct bicarbonate measurement with the new gem premier chemstat in hemodialysis patients. (1st September 2021)
- Record Type:
- Journal Article
- Title:
- Analytical evaluation of direct bicarbonate measurement with the new gem premier chemstat in hemodialysis patients. (1st September 2021)
- Main Title:
- Analytical evaluation of direct bicarbonate measurement with the new gem premier chemstat in hemodialysis patients
- Authors:
- Balboni, Fiamma
Terreni, Alessandro
Gallo, Marco
Guzzi, Francesco
Caparrini, Chiara
Burbui, Stefania
Vezzosi, Marco
Galora, Silvia
Lori, Giulia
Lippi, Giuseppe - Abstract:
- Abstract: GEM Premier ChemSTAT is a whole-blood analyzer designed for providing a rapid basic metabolic panel, inclusive of creatinine and blood urea nitrogen, with the unique characteristic of providing measured bicarbonate (HCO3 – ) levels. The aim of this work was to evaluate the clinical performance of HCO3 – assessment with this analyser in a real-life hemodialysis setting. Imprecision was calculated at different HCO3 – levels, along with assay comparison with Gem Premier 4000 analysers. GEM Premier ChemSTAT displayed an imprecision and a bias (in comparison to GEM Premier 4000) for HCO3 – of 0.4% and 37.3% at 20.8 mmol/L, 1.2% and 25.6% at 16.4 mmol/L, and 2.1% and 11.6% at 11.5 mmol/L, respectively, using three levels of HCO3 – quality control sample ChemSTAT System Evaluator. At direct comparison with the GEM Premier 4000 in the hemodialysis setting, Bland–Altman analysis of HCO3 – levels evidenced a bias ( µ ) of −4.9 (95% CI, −5.2 to −4.7) mmol/L. Such difference was attenuated by recalculating the GEM ChemSTAT expected HCO3 – values from pH and pCO2 using the Henderson Hasselbach equation, µ =–0.07 (95%CI, −0.19 to 0.05) mmol/L ( p = .24). In conclusion, our results show a remarkable difference between the HCO3 – values reported by GEM ChemSTAT or GEM 4000. New reference values for GEM ChemSTAT HCO3 – shall hence be defined according to our findings. We suggest that further investigation and a re-evaluation of the reference range should be made before extendingAbstract: GEM Premier ChemSTAT is a whole-blood analyzer designed for providing a rapid basic metabolic panel, inclusive of creatinine and blood urea nitrogen, with the unique characteristic of providing measured bicarbonate (HCO3 – ) levels. The aim of this work was to evaluate the clinical performance of HCO3 – assessment with this analyser in a real-life hemodialysis setting. Imprecision was calculated at different HCO3 – levels, along with assay comparison with Gem Premier 4000 analysers. GEM Premier ChemSTAT displayed an imprecision and a bias (in comparison to GEM Premier 4000) for HCO3 – of 0.4% and 37.3% at 20.8 mmol/L, 1.2% and 25.6% at 16.4 mmol/L, and 2.1% and 11.6% at 11.5 mmol/L, respectively, using three levels of HCO3 – quality control sample ChemSTAT System Evaluator. At direct comparison with the GEM Premier 4000 in the hemodialysis setting, Bland–Altman analysis of HCO3 – levels evidenced a bias ( µ ) of −4.9 (95% CI, −5.2 to −4.7) mmol/L. Such difference was attenuated by recalculating the GEM ChemSTAT expected HCO3 – values from pH and pCO2 using the Henderson Hasselbach equation, µ =–0.07 (95%CI, −0.19 to 0.05) mmol/L ( p = .24). In conclusion, our results show a remarkable difference between the HCO3 – values reported by GEM ChemSTAT or GEM 4000. New reference values for GEM ChemSTAT HCO3 – shall hence be defined according to our findings. We suggest that further investigation and a re-evaluation of the reference range should be made before extending the clinical use of this device. … (more)
- Is Part Of:
- Scandinavian journal of clinical & laboratory investigation. Volume 81:Number 5(2021)
- Journal:
- Scandinavian journal of clinical & laboratory investigation
- Issue:
- Volume 81:Number 5(2021)
- Issue Display:
- Volume 81, Issue 5 (2021)
- Year:
- 2021
- Volume:
- 81
- Issue:
- 5
- Issue Sort Value:
- 2021-0081-0005-0000
- Page Start:
- 418
- Page End:
- 421
- Publication Date:
- 2021-09-01
- Subjects:
- Performance -- measure error -- bed-side analysis -- potentiometry
Clinical biochemistry -- Periodicals
Physiology, Pathological -- Periodicals
Physiology, Experimental -- Periodicals
Medicine -- Research -- Periodicals
Clinical medicine -- Periodicals
616.0072 - Journal URLs:
- http://informahealthcare.com/loi/clb ↗
http://informahealthcare.com ↗ - DOI:
- 10.1080/00365513.2021.1930142 ↗
- Languages:
- English
- ISSNs:
- 0036-5513
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 8087.500000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 18512.xml