Right ventricular outflow tract stenting in symptomatic infants without the use of a long delivery sheath. Issue 2 (14th April 2021)
- Record Type:
- Journal Article
- Title:
- Right ventricular outflow tract stenting in symptomatic infants without the use of a long delivery sheath. Issue 2 (14th April 2021)
- Main Title:
- Right ventricular outflow tract stenting in symptomatic infants without the use of a long delivery sheath
- Authors:
- Linnane, Niall
Nasef, Mohamed Al
McMahon, Colin J.
McGuinness, Jonathan
McCrossan, Brian
Oslizlok, Paul
Walsh, Kevin P.
Kenny, Damien - Abstract:
- Abstract: Background and Objectives: Right ventricular outflow tract (RVOT) stenting improves systemic oxygenation and facilitates pulmonary arterial growth in symptomatic infants prior to repair of tetralogy of Fallot. The aim of this study was to evaluate the safety and efficacy of RVOT stenting without the use of a long delivery sheath. Methods: Retrospective data analysis of patients under 1 year of age undergoing RVOT stenting from January 2010 to January 2020 at a single tertiary pediatric cardiology center. Results: Sixty‐three RVOT stents were deployed during 53 procedures into 44 patients. The median age and weight at insertion were 41 days (range 2–204) and 3.6 kg (range 1.59–7) respectively. Thirty‐one procedures were semi‐elective and 22 were emergencies. Stent positioning was guided by transthoracic echocardiogram and/or RV angiography from a pigtail micro‐catheter placed via the aorta. The median total procedure and fluoroscopy times were 67.5 (range 15–145) and 19 min (1–107), respectively. The median length of hospital stay was 7 days (range 1–258). Twenty‐one patients were admitted to ICU post‐procedure with a median ICU length of stay of 3 days (range 3–11). There were three major complications including two deaths within 30 days of the procedure. A patient with Cornelia de Lange Syndrome (1.8 kg) died following stent migration and inability to wean from emergency cardiopulmonary bypass and the second infant had an unexplained asystolic arrestAbstract: Background and Objectives: Right ventricular outflow tract (RVOT) stenting improves systemic oxygenation and facilitates pulmonary arterial growth in symptomatic infants prior to repair of tetralogy of Fallot. The aim of this study was to evaluate the safety and efficacy of RVOT stenting without the use of a long delivery sheath. Methods: Retrospective data analysis of patients under 1 year of age undergoing RVOT stenting from January 2010 to January 2020 at a single tertiary pediatric cardiology center. Results: Sixty‐three RVOT stents were deployed during 53 procedures into 44 patients. The median age and weight at insertion were 41 days (range 2–204) and 3.6 kg (range 1.59–7) respectively. Thirty‐one procedures were semi‐elective and 22 were emergencies. Stent positioning was guided by transthoracic echocardiogram and/or RV angiography from a pigtail micro‐catheter placed via the aorta. The median total procedure and fluoroscopy times were 67.5 (range 15–145) and 19 min (1–107), respectively. The median length of hospital stay was 7 days (range 1–258). Twenty‐one patients were admitted to ICU post‐procedure with a median ICU length of stay of 3 days (range 3–11). There were three major complications including two deaths within 30 days of the procedure. A patient with Cornelia de Lange Syndrome (1.8 kg) died following stent migration and inability to wean from emergency cardiopulmonary bypass and the second infant had an unexplained asystolic arrest post‐procedure while awaiting transfer to ICU. Conclusions: RVOT stenting is technically possible with minimal complications without the need for a long delivery sheath. Additional imaging with transthoracic echocardiography can facilitate the safe deployment of the stent. … (more)
- Is Part Of:
- Catheterization and cardiovascular interventions. Volume 98:Issue 2(2021)
- Journal:
- Catheterization and cardiovascular interventions
- Issue:
- Volume 98:Issue 2(2021)
- Issue Display:
- Volume 98, Issue 2 (2021)
- Year:
- 2021
- Volume:
- 98
- Issue:
- 2
- Issue Sort Value:
- 2021-0098-0002-0000
- Page Start:
- E275
- Page End:
- E281
- Publication Date:
- 2021-04-14
- Subjects:
- congenital heart defects -- right ventricular outflow tract -- tetralogy of Fallot
Heart -- Diseases -- Diagnosis -- Periodicals
Cardiac catheterization -- Periodicals
616.1207572 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-726X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/ccd.29708 ↗
- Languages:
- English
- ISSNs:
- 1522-1946
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3092.992000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 18449.xml