Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement. Issue 2 (5th May 2021)
- Record Type:
- Journal Article
- Title:
- Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement. Issue 2 (5th May 2021)
- Main Title:
- Multicenter clinical evaluation of a piezoresistive‐MEMS‐sensor rapid‐exchange pressure microcatheter system for fractional flow reserve measurement
- Authors:
- Li, Chenguang
Yang, Junqing
Dong, Shaohong
Dong, Liang
Chen, Jiyan
Shen, Li
Zhang, Feng
Li, Changling
Liu, Huadong
Hu, Xinyang
Hau, William Kongto
Qian, Juying
Jeremias, Allen
Wang, Jian'an
Ge, Junbo - Abstract:
- Abstract: Objectives: This multicenter, prospective clinical study investigates whether the microelectromechanical‐systems‐(MEMS)‐sensor pressure microcatheter (MEMS‐PMC) is comparable to a conventional pressure wire in fractional flow reserve (FFR) measurement. Background: As a conventional tool for FFR measurement, pressure wires (PWs) still have some limitations such as suboptimal handling characteristics and unable to maintain the wire position during pullback assessment. Recently, a MEMS‐PMC compatible with any 0.014″ guidewire is developed. Compared with the existing optical‐sensor PMC, this MEMS‐PMC has smaller profiles at both the lesion crossing and sensor packaging areas. Methods: Two hundred and forty‐two patients with visually 30–70% coronary stenosis were enrolled at four centers. FFR was measured first with the MEMS‐PMC, and then with the PW. The primary endpoint was the Bland–Altman mean bias between the MEMS‐PMC and PW FFR. Results: From the 224‐patient per‐protocol data, quantitative coronary angiography showed 17.9% and 55.9% vessels had diameter < 2.5 mm and stenosis >50%, respectively. The two systems' mean bias was −0.01 with [−0.08, 0.06] 95% limits‐of‐agreement. Using PW FFR≤0.80 as cutoff, the MEMS‐PMC per‐vessel diagnostic accuracy was 93.4% [95% confidence interval: 89.4–96.3%]. The MEMS‐PMC's success rate was similar to that of PW (97.5 vs. 96.3%, p = .43) with no serious adverse event, and its clinically‐significant (>0.03) drift rate was 43%Abstract: Objectives: This multicenter, prospective clinical study investigates whether the microelectromechanical‐systems‐(MEMS)‐sensor pressure microcatheter (MEMS‐PMC) is comparable to a conventional pressure wire in fractional flow reserve (FFR) measurement. Background: As a conventional tool for FFR measurement, pressure wires (PWs) still have some limitations such as suboptimal handling characteristics and unable to maintain the wire position during pullback assessment. Recently, a MEMS‐PMC compatible with any 0.014″ guidewire is developed. Compared with the existing optical‐sensor PMC, this MEMS‐PMC has smaller profiles at both the lesion crossing and sensor packaging areas. Methods: Two hundred and forty‐two patients with visually 30–70% coronary stenosis were enrolled at four centers. FFR was measured first with the MEMS‐PMC, and then with the PW. The primary endpoint was the Bland–Altman mean bias between the MEMS‐PMC and PW FFR. Results: From the 224‐patient per‐protocol data, quantitative coronary angiography showed 17.9% and 55.9% vessels had diameter < 2.5 mm and stenosis >50%, respectively. The two systems' mean bias was −0.01 with [−0.08, 0.06] 95% limits‐of‐agreement. Using PW FFR≤0.80 as cutoff, the MEMS‐PMC per‐vessel diagnostic accuracy was 93.4% [95% confidence interval: 89.4–96.3%]. The MEMS‐PMC's success rate was similar to that of PW (97.5 vs. 96.3%, p = .43) with no serious adverse event, and its clinically‐significant (>0.03) drift rate was 43% less (9.5 vs. 16.7%, p = .014). Conclusions: Our study showed the MEMS‐PMC is safe to use and has a minimal bias equal to the resolution of current FFR systems. Given the MEMS‐PMC's high measurement accuracy and rapid‐exchange nature, it may become an attractive new tool facilitating routine coronary physiology assessment. … (more)
- Is Part Of:
- Catheterization and cardiovascular interventions. Volume 98:Issue 2(2021)
- Journal:
- Catheterization and cardiovascular interventions
- Issue:
- Volume 98:Issue 2(2021)
- Issue Display:
- Volume 98, Issue 2 (2021)
- Year:
- 2021
- Volume:
- 98
- Issue:
- 2
- Issue Sort Value:
- 2021-0098-0002-0000
- Page Start:
- E243
- Page End:
- E253
- Publication Date:
- 2021-05-05
- Subjects:
- coronary artery disease -- catheterization -- diagnostic -- clinical trials -- coronary blood flow/physiology/microvascular function -- fractional flow reserve -- percutaneous coronary intervention
Heart -- Diseases -- Diagnosis -- Periodicals
Cardiac catheterization -- Periodicals
616.1207572 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-726X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/ccd.29678 ↗
- Languages:
- English
- ISSNs:
- 1522-1946
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3092.992000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 18437.xml