In-line Pullout Strength of 2 Acetabular Fixation Methods for Ligamentum Teres Reconstruction of the Hip: A Cadaveric Study. Issue 12 (2nd December 2021)
- Record Type:
- Journal Article
- Title:
- In-line Pullout Strength of 2 Acetabular Fixation Methods for Ligamentum Teres Reconstruction of the Hip: A Cadaveric Study. Issue 12 (2nd December 2021)
- Main Title:
- In-line Pullout Strength of 2 Acetabular Fixation Methods for Ligamentum Teres Reconstruction of the Hip: A Cadaveric Study
- Authors:
- Lall, Ajay C.
Ankem, Hari K.
Ryan, Michael K.
Beason, David P.
Diulus, Samantha C.
Roach, Ryan P.
Rosinsky, Philip J.
Maldonado, David R.
Emblom, Benton A.
Domb, Benjamin G. - Abstract:
- Background: Ligamentum teres (LT) reconstruction is an appropriate alternative in select cases of LT full-thickness tears, resulting in hip micro- or macroinstability. Graft fixation at the acetabular fossa is critical to achieving the best functional results. Purpose: The purpose of this study is to compare the pullout strength of 2 graft fixation methods used for LT reconstruction of the hip. Study Design: Controlled laboratory study. Methods: In 7 cadaveric specimens, the acetabular socket was prepared after the native LT was transected and the femoral head was removed. Seven separate tibialis anterior grafts were then prepared by suturing a running-locking No. 2 suture on each tail of the graft. Three specimens had fixation of the graft to the acetabulum using an adjustable cortical suspension suture button; the remaining 4 were fixed to the acetabulum using a knotless suture anchor. Specimens were then mounted onto a custom jig within a mechanical test frame to allow for the in-line pull of the graft fixation construct. After a preload of 5 N, each specimen was loaded to failure at 0.5 mm/s. Stiffness and load to failure were measured for each specimen construct. Results: Suture button fixation had a higher mean load to failure when compared with the knotless anchor fixation method (mean ± SD, 438.1 ± 114.3 vs 195.9 ± 50.0 N; P = .01). There was no significant difference in mean stiffness between the methods of fixation (24.5 ± 1.4 vs 26.5 ± 5.8 N/mm; P = .6).Background: Ligamentum teres (LT) reconstruction is an appropriate alternative in select cases of LT full-thickness tears, resulting in hip micro- or macroinstability. Graft fixation at the acetabular fossa is critical to achieving the best functional results. Purpose: The purpose of this study is to compare the pullout strength of 2 graft fixation methods used for LT reconstruction of the hip. Study Design: Controlled laboratory study. Methods: In 7 cadaveric specimens, the acetabular socket was prepared after the native LT was transected and the femoral head was removed. Seven separate tibialis anterior grafts were then prepared by suturing a running-locking No. 2 suture on each tail of the graft. Three specimens had fixation of the graft to the acetabulum using an adjustable cortical suspension suture button; the remaining 4 were fixed to the acetabulum using a knotless suture anchor. Specimens were then mounted onto a custom jig within a mechanical test frame to allow for the in-line pull of the graft fixation construct. After a preload of 5 N, each specimen was loaded to failure at 0.5 mm/s. Stiffness and load to failure were measured for each specimen construct. Results: Suture button fixation had a higher mean load to failure when compared with the knotless anchor fixation method (mean ± SD, 438.1 ± 114.3 vs 195.9 ± 50.0 N; P = .01). There was no significant difference in mean stiffness between the methods of fixation (24.5 ± 1.4 vs 26.5 ± 5.8 N/mm; P = .6). Conclusion: In this cadaveric study, the suture button fixation demonstrated greater load to failure than the knotless anchor fixation. Clinical Relevance: Results of this study can guide surgical decision making when selecting an acetabular fixation method for LT reconstruction. … (more)
- Is Part Of:
- Orthopaedic journal of sports medicine. Volume 9:Issue 12(2021)
- Journal:
- Orthopaedic journal of sports medicine
- Issue:
- Volume 9:Issue 12(2021)
- Issue Display:
- Volume 9, Issue 12 (2021)
- Year:
- 2021
- Volume:
- 9
- Issue:
- 12
- Issue Sort Value:
- 2021-0009-0012-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-12-02
- Subjects:
- ligamentum teres reconstruction -- pullout strength -- hip arthroscopy -- anchor -- suture button
Sports medicine -- Periodicals
Orthopedics -- Periodicals
Arthroscopy -- Periodicals
Arthroplasty -- Periodicals
Knee -- Surgery -- Periodicals
616.7 - Journal URLs:
- http://www.sagepublications.com/ ↗
- DOI:
- 10.1177/23259671211052533 ↗
- Languages:
- English
- ISSNs:
- 2325-9671
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 18403.xml