SAT0560 Automated Ifa Methods Compare Well with Established Manual IFA Screening and Titration for ANA HEp-2. (15th July 2016)
- Record Type:
- Journal Article
- Title:
- SAT0560 Automated Ifa Methods Compare Well with Established Manual IFA Screening and Titration for ANA HEp-2. (15th July 2016)
- Main Title:
- SAT0560 Automated Ifa Methods Compare Well with Established Manual IFA Screening and Titration for ANA HEp-2
- Authors:
- Blecken, J.
Matthias, T.
Frey, A.
Daves, M.
Joos, A.
Platzgummer, S. - Abstract:
- Abstract : Background: Often, only basic nuclear patterns, like homogenous, speckled, nucleolar, centromere or cytoplasmic, are reported by laboratories. The detection of other patterns requires well trained readers. As a result, different systems have been developed which automate part of or the complete IFA method and reading process. Objectives: This study compares 2 commercially available HEp-2 antinuclear antibody (ANA) indirect fluorescent antibody (IFA) assays using a sensitivity panel (120 clinically determined patients) and a specificity panel consisting of 80 clinically confirmed negative patients. Methods: We compared two different automated systems (A and B) to assess their capability for screening and titration of these samples. The automated method was directly compared to manual reading of the same processed slides on respective microscopes and also compared with the known clinical information. Results: The results of the two automated methods were in good agreement. System A detected 188 samples correctly from negative and positive samples, versus 187 samples detected by the system B. The falsely detected positive samples were all of low titer (1:80). System A found 157 and System B 156 patterns in agreement to the target pattern. From 80 negative samples system A detected 73 and system B 71 samples correctly. Conclusions: Both systems resulted in an overall sensitivity >95% and a specificity of 91.25% and 88.75%. The pattern recognition also showed onlyAbstract : Background: Often, only basic nuclear patterns, like homogenous, speckled, nucleolar, centromere or cytoplasmic, are reported by laboratories. The detection of other patterns requires well trained readers. As a result, different systems have been developed which automate part of or the complete IFA method and reading process. Objectives: This study compares 2 commercially available HEp-2 antinuclear antibody (ANA) indirect fluorescent antibody (IFA) assays using a sensitivity panel (120 clinically determined patients) and a specificity panel consisting of 80 clinically confirmed negative patients. Methods: We compared two different automated systems (A and B) to assess their capability for screening and titration of these samples. The automated method was directly compared to manual reading of the same processed slides on respective microscopes and also compared with the known clinical information. Results: The results of the two automated methods were in good agreement. System A detected 188 samples correctly from negative and positive samples, versus 187 samples detected by the system B. The falsely detected positive samples were all of low titer (1:80). System A found 157 and System B 156 patterns in agreement to the target pattern. From 80 negative samples system A detected 73 and system B 71 samples correctly. Conclusions: Both systems resulted in an overall sensitivity >95% and a specificity of 91.25% and 88.75%. The pattern recognition also showed only minor aberrant findings resulting in a slightly better detection of cytoplasmic and nuclear membrane patterns by system A while system B detected slightly better the centromeric pattern. Disclosure of Interest: None declared … (more)
- Is Part Of:
- Annals of the rheumatic diseases. Volume 75(2016)Supplement 2
- Journal:
- Annals of the rheumatic diseases
- Issue:
- Volume 75(2016)Supplement 2
- Issue Display:
- Volume 75, Issue 2 (2016)
- Year:
- 2016
- Volume:
- 75
- Issue:
- 2
- Issue Sort Value:
- 2016-0075-0002-0000
- Page Start:
- 872
- Page End:
- 872
- Publication Date:
- 2016-07-15
- Subjects:
- Rheumatism -- Periodicals
616.723005 - Journal URLs:
- http://ard.bmjjournals.com/ ↗
http://www.pubmedcentral.nih.gov/tocrender.fcgi?journal=149&action=archive ↗
http://www.bmj.com/archive ↗
http://gateway.ovid.com/server3/ovidweb.cgi?T=JS&MODE=ovid&D=ovft&PAGE=titles&SEARCH=annals+of+the+rheumatic+diseases.tj&NEWS=N ↗ - DOI:
- 10.1136/annrheumdis-2016-eular.3958 ↗
- Languages:
- English
- ISSNs:
- 0003-4967
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 18356.xml