Manipulative and inherent errors in anti-D quantitation using the AutoAnalyzer. Issue 3 (March 1972)
- Record Type:
- Journal Article
- Title:
- Manipulative and inherent errors in anti-D quantitation using the AutoAnalyzer. Issue 3 (March 1972)
- Main Title:
- Manipulative and inherent errors in anti-D quantitation using the AutoAnalyzer
- Authors:
- Gunson, H. H.
Phillips, P. K.
Stratton, F. - Abstract:
- Abstract : The limits of accuracy for the quantitation of anti-D using the AutoAnalyzer have been reduced from ± 26% with manually prepared dilutions to ± 14% using an automatic pipette, as expressed by the 95% confidence limits. The error inherent in the AutoAnalyzer was estimated to contribute ± 10% to the overall error. Problems associated with the reproducibility of this method for anti-D quantitation have been investigated, namely, the effect of the age of a given test cell, the use of different test cells, loss of sensitivity of the machine over a given day, and the reproducibility of results obtained at different positions on the standard graph.
- Is Part Of:
- Journal of clinical pathology. Volume 25:Issue 3(1972)
- Journal:
- Journal of clinical pathology
- Issue:
- Volume 25:Issue 3(1972)
- Issue Display:
- Volume 25, Issue 3 (1972)
- Year:
- 1972
- Volume:
- 25
- Issue:
- 3
- Issue Sort Value:
- 1972-0025-0003-0000
- Page Start:
- 198
- Page End:
- 205
- Publication Date:
- 1972-03
- Subjects:
- Pathology -- Periodicals
Pathology, Molecular -- Periodicals
616.0705 - Journal URLs:
- http://jcp.bmjjournals.com ↗
http://jcp.bmjjournals.com/content/by/year ↗
http://www.pubmedcentral.nih.gov/tocrender.fcgi?journal=162&action=archive ↗
http://www.bmj.com/archive ↗ - DOI:
- 10.1136/jcp.25.3.198 ↗
- Languages:
- English
- ISSNs:
- 0021-9746
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 18343.xml