Full core technology versus notch sampling technology: evaluation of the diagnostic accuracy and the risk of a pneumothorax after transthoracic needle biopsy of suspicious lung lesions. (January 2022)
- Record Type:
- Journal Article
- Title:
- Full core technology versus notch sampling technology: evaluation of the diagnostic accuracy and the risk of a pneumothorax after transthoracic needle biopsy of suspicious lung lesions. (January 2022)
- Main Title:
- Full core technology versus notch sampling technology: evaluation of the diagnostic accuracy and the risk of a pneumothorax after transthoracic needle biopsy of suspicious lung lesions
- Authors:
- Schroeder, Christophe
Loebelenz, Laura I
Heverhagen, Johannes T
Noeldge, Gerd
Brönnimann, Michael P
Kim, Soung Yung - Abstract:
- Background: Percutaneous needle biopsy of the lung (PCBL) under image guidance has become a safe and effective minimal invasive method to obtain a specimen related histological diagnosis of pulmonary lesions. Purpose: To evaluate the diagnostic yield and safety of two different coaxial biopsy technologies: full core and notch sampling technology. The former allowing the removal of full punch cylinders and the latter using a cutting-edge mechanism. Material and Methods: A retrospective analysis of 48 consecutive PCBL procedures has been carried out for this prognostic study, involving patients with a documented pulmonary nodule or mass lesion on previous computed tomography (CT) scans. The study population included 38 men and 10 women (mean age = 67 years). Of these 48 patients who underwent a procedure with a co-axial cutting system, 24 have been performed with notch sampling technology and 24 with full core technology. Results: Out of the 48 biopsy procedures, 46 yielded specimens were adequate for histopathological evaluation, consistent with a technical success rate of 96%. The most common induced image-guided biopsy complication was a pneumothorax, occurring in 14 patients (35%). Seven patients with a pneumothorax were attributed to the full core technology and seven to the notch sampling technology (odds ratio = 1, 95% confidence interval = 0.28–3.51, P = 1). Conclusion: In the setting of full core versus notch sampling percutaneous CT-guided coaxial needle biopsy ofBackground: Percutaneous needle biopsy of the lung (PCBL) under image guidance has become a safe and effective minimal invasive method to obtain a specimen related histological diagnosis of pulmonary lesions. Purpose: To evaluate the diagnostic yield and safety of two different coaxial biopsy technologies: full core and notch sampling technology. The former allowing the removal of full punch cylinders and the latter using a cutting-edge mechanism. Material and Methods: A retrospective analysis of 48 consecutive PCBL procedures has been carried out for this prognostic study, involving patients with a documented pulmonary nodule or mass lesion on previous computed tomography (CT) scans. The study population included 38 men and 10 women (mean age = 67 years). Of these 48 patients who underwent a procedure with a co-axial cutting system, 24 have been performed with notch sampling technology and 24 with full core technology. Results: Out of the 48 biopsy procedures, 46 yielded specimens were adequate for histopathological evaluation, consistent with a technical success rate of 96%. The most common induced image-guided biopsy complication was a pneumothorax, occurring in 14 patients (35%). Seven patients with a pneumothorax were attributed to the full core technology and seven to the notch sampling technology (odds ratio = 1, 95% confidence interval = 0.28–3.51, P = 1). Conclusion: In the setting of full core versus notch sampling percutaneous CT-guided coaxial needle biopsy of the lung, no significant difference in the diagnostic accuracy and the incidence of pneumothoraces could be shown, while both technologies have an excellent diagnostic performance. … (more)
- Is Part Of:
- Acta radiologica. Volume 63:Number 1(2022)
- Journal:
- Acta radiologica
- Issue:
- Volume 63:Number 1(2022)
- Issue Display:
- Volume 63, Issue 1 (2022)
- Year:
- 2022
- Volume:
- 63
- Issue:
- 1
- Issue Sort Value:
- 2022-0063-0001-0000
- Page Start:
- 35
- Page End:
- 41
- Publication Date:
- 2022-01
- Subjects:
- Percutaneous lung biopsy -- image guidance -- pulmonary lesions -- full core technology -- diagnostic accuracy -- notch sampling technology -- efficacy -- safety
Radiology, Medical -- Periodicals
Radiography, Medical -- Periodicals
Radiotherapy -- Periodicals
616.0757 - Journal URLs:
- http://acr.sagepub.com ↗
http://ar.rsmjournals.com ↗
http://www.uk.sagepub.com/home.nav ↗
http://informahealthcare.com/loi/ard ↗
http://www.tandf.co.uk/journals/titles/02841851.asp ↗ - DOI:
- 10.1177/0284185120981575 ↗
- Languages:
- English
- ISSNs:
- 0284-1851
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0662.000000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 18237.xml