An automatic incision device for obtaining blood samples from the heels of preterm infants causes less damage than a conventional manual lancet. Issue 1 (1st January 2001)
- Record Type:
- Journal Article
- Title:
- An automatic incision device for obtaining blood samples from the heels of preterm infants causes less damage than a conventional manual lancet. Issue 1 (1st January 2001)
- Main Title:
- An automatic incision device for obtaining blood samples from the heels of preterm infants causes less damage than a conventional manual lancet
- Authors:
- Vertanen, H
Fellman, V
Brommels, M
Viinikka, L - Abstract:
- Abstract : OBJECTIVES: To evaluate in a randomised blind study the effect on puncture site lesions of two different incision devices used to obtain blood samples from preterm infants by repeated heel sticks. SETTING: The neonatal intensive care unit at the Hospital for Children and Adolescents and Laboratory, Helsinki University Central Hospital. PATIENTS: A total of 100 preterm infants (birth weight below 2500 g) not previously subjected to heel stick sampling. INTERVENTIONS: The infants were randomly allocated to blood sampling from the heel with either a conventional manual lancet or an automatic incision device. The same type of lancet was used for any given baby throughout the study (2–21 days). MAIN OUTCOME MEASURES: The damage caused by sampling was evaluated using four criteria: bruising of the heel, inflammation of the heel, bruising of either the ankle or the leg, and skin healing at the puncture site. The evaluation was based on photographs presenting typical categories of each outcome. RESULTS: To obtain a sufficient volume of blood, on average 2.6 times more punctures were needed when the conventional manual lancet was used than when the automatic incision device was used. Heels punctured with the lancet had more bruising (100% v 84%) and more signs of inflammation (79% v 53%), and there was more bruising of the ankle or leg (92% v 53%) than when the automatic incision device was used. Skin healed equally rapidly in the two groups. CONCLUSION: The use of anAbstract : OBJECTIVES: To evaluate in a randomised blind study the effect on puncture site lesions of two different incision devices used to obtain blood samples from preterm infants by repeated heel sticks. SETTING: The neonatal intensive care unit at the Hospital for Children and Adolescents and Laboratory, Helsinki University Central Hospital. PATIENTS: A total of 100 preterm infants (birth weight below 2500 g) not previously subjected to heel stick sampling. INTERVENTIONS: The infants were randomly allocated to blood sampling from the heel with either a conventional manual lancet or an automatic incision device. The same type of lancet was used for any given baby throughout the study (2–21 days). MAIN OUTCOME MEASURES: The damage caused by sampling was evaluated using four criteria: bruising of the heel, inflammation of the heel, bruising of either the ankle or the leg, and skin healing at the puncture site. The evaluation was based on photographs presenting typical categories of each outcome. RESULTS: To obtain a sufficient volume of blood, on average 2.6 times more punctures were needed when the conventional manual lancet was used than when the automatic incision device was used. Heels punctured with the lancet had more bruising (100% v 84%) and more signs of inflammation (79% v 53%), and there was more bruising of the ankle or leg (92% v 53%) than when the automatic incision device was used. Skin healed equally rapidly in the two groups. CONCLUSION: The use of an automatic incision device for collecting repeated skin puncture samples from preterm infants is less traumatic than the use of a conventional manual lancet. … (more)
- Is Part Of:
- Archives of disease in childhood. Volume 84:Issue 1(2001)
- Journal:
- Archives of disease in childhood
- Issue:
- Volume 84:Issue 1(2001)
- Issue Display:
- Volume 84, Issue 1 (2001)
- Year:
- 2001
- Volume:
- 84
- Issue:
- 1
- Issue Sort Value:
- 2001-0084-0001-0000
- Page Start:
- F53
- Page End:
- F55
- Publication Date:
- 2001-01-01
- Subjects:
- newborn -- blood sampling -- skin puncture -- puncture site healing -- heel
Infants -- Diseases -- Periodicals
Newborn infants -- Diseases -- Periodicals
Fetus -- Diseases -- Periodicals
618.920105 - Journal URLs:
- http://fn.bmjjournals.com ↗
http://www.bmj.com/archive ↗ - DOI:
- 10.1136/fn.84.1.F53 ↗
- Languages:
- English
- ISSNs:
- 1359-2998
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 18190.xml