False negative findings in intraoperative SEP monitoring: analysis of 658 consecutive neurosurgical cases and review of published reports. Issue 2 (23rd January 2004)
- Record Type:
- Journal Article
- Title:
- False negative findings in intraoperative SEP monitoring: analysis of 658 consecutive neurosurgical cases and review of published reports. Issue 2 (23rd January 2004)
- Main Title:
- False negative findings in intraoperative SEP monitoring: analysis of 658 consecutive neurosurgical cases and review of published reports
- Authors:
- Wiedemayer, H
Sandalcioglu, I E
Armbruster, W
Regel, J
Schaefer, H
Stolke, D - Abstract:
- Abstract : Objectives: To determine the sensitivity of intraoperative monitoring in neurosurgical operations using somatosensory evoked potentials and to identify reasons for false negative findings and possible settings with an increased risk for monitoring failure. Methods: SEP monitoring of 658 neurosurgical operations was analysed. The target of monitoring was the function of a hemisphere in 251 cases, the brain stem in 198 cases, and the spinal cord in 209 cases. Results: In 27 cases (4.1%), monitoring was classified as false negative. Further analysis showed that five of these patients had experienced delayed neurological damage. Among the remaining 22 false negative cases, 14 had a minor neurological deficit and eight had severe neurological damage. Overall sensitivity and negative predictive value of SEP monitoring was 79% and 96%, respectively. For the detection of severe neurological damage the corresponding figures were 91% and 98%. Sensitivity of monitoring varied depending on the target of monitoring and the type of lesion. Monitoring was less likely to detect neurological damage in surgery for infratentorial tumours with brain stem compression, small lesions of the motor cortex, and small vessel damage during aneurysm surgery. Conclusions: SEP monitoring has acceptable sensitivity for detecting neurological damage during different neurosurgical procedures. Distinct settings with an increased risk of monitoring failure can be identified. In these cases measuresAbstract : Objectives: To determine the sensitivity of intraoperative monitoring in neurosurgical operations using somatosensory evoked potentials and to identify reasons for false negative findings and possible settings with an increased risk for monitoring failure. Methods: SEP monitoring of 658 neurosurgical operations was analysed. The target of monitoring was the function of a hemisphere in 251 cases, the brain stem in 198 cases, and the spinal cord in 209 cases. Results: In 27 cases (4.1%), monitoring was classified as false negative. Further analysis showed that five of these patients had experienced delayed neurological damage. Among the remaining 22 false negative cases, 14 had a minor neurological deficit and eight had severe neurological damage. Overall sensitivity and negative predictive value of SEP monitoring was 79% and 96%, respectively. For the detection of severe neurological damage the corresponding figures were 91% and 98%. Sensitivity of monitoring varied depending on the target of monitoring and the type of lesion. Monitoring was less likely to detect neurological damage in surgery for infratentorial tumours with brain stem compression, small lesions of the motor cortex, and small vessel damage during aneurysm surgery. Conclusions: SEP monitoring has acceptable sensitivity for detecting neurological damage during different neurosurgical procedures. Distinct settings with an increased risk of monitoring failure can be identified. In these cases measures to enhance the sensitivity of monitoring should be considered. … (more)
- Is Part Of:
- Journal of neurology, neurosurgery and psychiatry. Volume 75:Issue 2(2004)
- Journal:
- Journal of neurology, neurosurgery and psychiatry
- Issue:
- Volume 75:Issue 2(2004)
- Issue Display:
- Volume 75, Issue 2 (2004)
- Year:
- 2004
- Volume:
- 75
- Issue:
- 2
- Issue Sort Value:
- 2004-0075-0002-0000
- Page Start:
- 280
- Page End:
- 286
- Publication Date:
- 2004-01-23
- Subjects:
- SEP monitoring -- neurosurgery
BAEP, brain stem auditory evoked potential -- IOM, intraoperative neurophysiological monitoring -- M-SEP, median nerve somatosensory evoked potential -- SEP, somatosensory evoked potential -- T-SEP, tibial nerve somatosensory evoked potential
Neurology -- Periodicals
Nervous system -- Surgery -- Periodicals
Psychiatry -- Periodicals
616.8 - Journal URLs:
- http://jnnp.bmjjournals.com/ ↗
http://www.pubmedcentral.nih.gov/tocrender.fcgi?action=archive&journal=192 ↗
http://www.bmj.com/archive ↗ - DOI:
- 10.1136/jnnp.2002.008904 ↗
- Languages:
- English
- ISSNs:
- 0022-3050
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 18038.xml